Key result
Restoring pulsatility modulates shear-induced von Willebrand factor degradation during continuous-flow mechanical circulatory support.
Why the study?
The relationship between arterial pulsatility and the acquired von Willebrand factor defect in patients with continuous-flow mechanical circulatory support had not been studied.
Does restoring arterial pulsatility improve von Willebrand factor (VWF) multimer degradation during continuous-flow mechanical circulatory support?
Does restoring arterial pulsatility improve von Willebrand factor (VWF) multimer degradation during continuous-flow mechanical circulatory support?
Restoration of arterial pulsatility mitigates the acquired von Willebrand factor defect in continuous-flow mechanical circulatory support, providing a potential mechanism and target to reduce bleeding complications.
Background The main risk factor for bleeding in patients with continuous-flow mechanical circulatory support (CF-MCS) is the acquired von Willebrand factor (VWF) defect related to the high shear-stress forces developed by these devices. Although a higher bleeding rate has been reported in CF-MCS recipients who had reduced pulsatility, the relation between pulsatility and the VWF defect has never been studied. Objectives The purpose of this study was to investigate the relation between pulsatility and VWF under CF-MCS. Methods We assessed the effect of 2 CF-MCS on VWF multimer degradation in a mock circulatory loop (model 1). Using these devices, we investigated in a dose-effect model (model 2) 3 levels of pulsatility in 3 groups of swine. In a cross-over model (model 3), we studied the effects of sequential changes of pulsatility on VWF. We reported the evolution of VWF multimerization in a patient undergoing serial CF-MCS and/or pulsatile-MCS. Results We demonstrated the proteolytic degradation of VWF multimers by high shear CF-MCS in a circulatory loop without pulsatility. We observed both in swine models and in a patient that the magnitude of the VWF degradation is modulated by the pulsatility level in the high shear-stress level condition, and that the restoration of pulsatility is a trigger for the endothelial release of VWF. Conclusions We demonstrated that the VWF defect reflects the balance between degradation induced by the shear stress and the endothelial release of new VWF triggered by the pulsatility. This modulation of VWF levels could explain the relationship between pulsatility and bleeding observed in CF-MCS recipients. Preservation of pulsatility may be a new target to improve clinical outcomes of patients.
No takes yet. Share an insight, caveat, or question.
Vincent et al. (2018) studied Continuous-flow mechanical circulatory support. Arterial pulsatility vs. Continuous-flow mechanical circulatory support without pulsatility was evaluated on von Willebrand factor (VWF) multimer degradation. In experimental swine models and a single patient, the magnitude of von Willebrand factor degradation by high shear-stress continuous-flow mechanical circulatory support was modulated by pulsatility.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: