Key result
Novel pressure wire records ~27 mmHg transstenotic gradient reduction after coronary angioplasty.
Why the study?
Measurements of the transstenotic pressure gradient are advocated to assess coronary angioplasty results but are unreliable with angioplasty catheters.
Does a novel 0.015-in. fluid-filled guide wire accurately measure transstenotic pressure gradients compared to actual pressure and quantitative coronary angiography?
Observational (n=37)
Does a novel 0.015-in. fluid-filled guide wire accurately measure transstenotic pressure gradients compared to actual pressure and quantitative coronary angiography?
Answers are grounded in this paper and cite their sources.
A 0.015-in. fluid-filled guide wire can accurately and reliably measure mean transstenotic pressure gradients, facilitating functional assessment of coronary stenoses and postangioplasty results.
No abstract yet. The journal hasn't published one we can show.
No takes yet. Share an insight, caveat, or question.
Bruyne et al. (1993) conducted an observational in Coronary stenosis (n=37). 0.015-in. fluid-filled guide wire for pressure measurement vs. Before angioplasty was evaluated on Mean transstenotic pressure gradient (p=< 0.01). Mean transstenotic pressure gradients recorded with a novel 0.015-in. fluid-filled guide wire decreased from 30 mm Hg before to 3 mm Hg after coronary angioplasty (p<0.01).