Key result
Pulsed field ablation linked to ~17-minute shorter procedure times than cryoballoon in de novo ablation.
Why the study?
Efficient use of EP laboratory resources requires accurate prediction of procedural durations for AF ablation, and it is unclear if PFA improves procedural time predictability compared with thermal ablation.
Does pulsed field ablation improve procedural time predictability and reduce EP lab time compared with thermal ablation in real-world de novo and redo atrial fibrillation ablation?
Observational (n=1,845)
Does pulsed field ablation improve procedural time predictability and reduce EP lab time compared with thermal ablation in real-world de novo and redo atrial fibrillation ablation?
Pulsed field ablation for atrial fibrillation yields significantly shorter and more predictable procedural times compared to thermal ablation, potentially improving electrophysiology laboratory workflow and efficiency.
Background Efficient use of electrophysiology (EP) laboratory resources is enhanced by accurate prediction of procedural durations for atrial fibrillation (AF) ablation. The availability of devices with reproducible procedural durability further supports this process, improving scheduling efficiency and overall resource utilization. Objective This study evaluated whether pulsed field ablation (PFA) improves procedural time predictability compared with thermal ablation (radiofrequency (RF) and cryoablation (CRYO)) in real-world de novo and redo AF ablation. Methods Retrospective analysis comparing procedural variability using Levene, Brown–Forsythe, and trimmed-mean tests. Determinants of EP lab time were identified via multivariable regression with log-transformed time as the dependent variable. Results A total of 1845 procedures were analyzed, comprising 1730 de novo cases (599 PFA-Pulmonary Vein Isolation [PVI], 531 PFA-PVI+Posterior Wall Isolation [PWI], 600 CRYO-PVI) and 115 redo cases (69 PFA, 46 RF). In de-novo procedures, mean procedural time was shorter with PFA-PVI (58.5 ± 23.4 min) than with PFA-PVI+PWI (71.6 ± 31.1 min) and cryoballoon PVI (75.7 ± 39.5 min; p<0.001). Corresponding EP lab times were 94.5 ± 28.7, 110.1 ± 37.2, and 109.3 ± 41.7 min (p<0.001). In redo procedures, PFA required significantly less EP lab time than RF (102.5 ± 22.0 vs. 156.8 ± 49.5 min; p<0.001). Across all variance tests, PFA demonstrated significantly lower variability in both de novo and redo cases (p<0.001). Multivariable analysis confirmed PFA reduced EP lab time in de novo (β = –0.142) and redo procedures (β =-0.340; both p<0.001).Factors associated with longer procedural times included PWI, cavotricuspid isthmus ablation, presence of cardiac implantable electronic devices, and prior cardiac surgery. Conclusions PFA delivers more predictable procedural times than thermal ablation. Integrating procedural time predictability into ablation modality selection may enhance workflow efficiency, optimize EP lab scheduling, and improve overall procedural throughput.
No takes yet. Share an insight, caveat, or question.
Awada et al. (2026) conducted an observational in Paroxysmal and persistent atrial fibrillation (n=1,845). Pulsed field ablation (PFA) vs. Thermal ablation (radiofrequency and cryoablation) was evaluated on Mean procedural time in de novo ablation (PFA-PVI vs cryoballoon PVI, in minutes) (p=<0.001). Pulsed field ablation resulted in significantly shorter and more predictable procedural times compared to cryoballoon ablation in de novo procedures (58.5 vs 75.7 minutes; P<0.001).