Key result
Bioprosthetic valve fracture during ViV TAVR reduces mean transvalvular gradient by ~37 mmHg.
Why the study?
Bioprosthetic valve fracturing during valve-in-valve TAVR may reduce transvalvular gradients and increase aortic valve area in patients with failed bioprosthetic valves, but data are limited.
Cohort (n=12)
No
Transcatheter aortic valve replacement (TAVR) valve-in-valve (VIV) therapy has been approved for select patients with surgically inoperable bioprosthetic valves that need replacement. Bioprosthetic valve fracturing (BVF) used in conjunction with VIV TAVR can reduce transvalvular gradients and increase the aortic valve area. Twelve patients who underwent BVF VIV TAVR at a single center were retrospectively analyzed. Measurements of hemodynamics and aortic valve area were performed at baseline, after VIV TAVR, after BVF, and at 30-day follow-up. The mean Society of Thoracic Surgeons Predicted Risk of Mortality score was 7.12 ± 3.5%, with 75% of patients deemed high risk by the heart team. Mean gradients decreased from 44 mm Hg to 15 mm Hg following VIV TAVR, and to 7 mm Hg following BVF. The mean aortic valve area increased from 0.6 cm2 to 1 cm2 following VIV TAVR, and to 1.3 cm2 following BVF. There were no postoperative permanent pacemaker implantations or vascular complications, and at 30 days, only one patient had died. While we report intraoperative mortality, BVF with VIV TAVR can be performed to reduce transvalvular gradients and increase effective aortic valve area in high-surgical-risk patients with failed bioprosthetic valves.
No takes yet. Share an insight, caveat, or question.
Hamandi et al. (2020) conducted a cohort in Failed surgically inoperable bioprosthetic valves (n=12). Bioprosthetic valve fracturing during valve-in-valve TAVR vs. Baseline (pre-procedure) was evaluated on Mean transvalvular gradient and aortic valve area. Bioprosthetic valve fracturing with valve-in-valve TAVR reduced mean transvalvular gradients from 44 mm Hg at baseline to 7 mm Hg, and increased mean aortic valve area from 0.6 cm2 to 1.3 cm2.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: