Key result
Screw-in temporary pacing leads linked to ~76-point lower ICU utilization vs conventional leads after TAVR.
Why the study?
Patients with pre-existing right bundle branch block undergoing TAVR are at high risk of complete heart block requiring prolonged temporary pacing, but conventional leads have limitations such as lead dislodgement and need for ICU monitoring.
Does the use of screw-in temporary pacing leads reduce ICU utilization and hospital length of stay in TAVR patients with pre-existing RBBB?
Observational (n=544)
Yes
Does the use of screw-in temporary pacing leads reduce ICU utilization and hospital length of stay in TAVR patients with pre-existing RBBB?
Absolute Event Rate: 23% vs 99%
p-value: p=<0.001
In TAVR patients with pre-existing RBBB, using screw-in temporary pacing leads reduces ICU utilization and shortens hospital stay compared to conventional leads without increasing complications.
Screw-in leads were associated with lower ICU use after TAVR in RBBB patients; hypothesis-generating and requires randomized confirmation before practice change.
STRUCTURED ABSTRACT Background Patients with pre-existing right bundle branch block (RBBB) undergoing transcatheter aortic valve replacement (TAVR) are at high risk of complete heart block, often requiring prolonged temporary pacing. Conventional temporary pacing leads are associated with lead dislodgement, immobilization and intensive care unit (ICU) monitoring. The objective of this study was to compare screw-in versus conventional temporary pacing leads after TAVR in patients with pre-existing RBBB. Methods This multicenter observational study included consecutive patients with pre-existing RBBB undergoing TAVR between 2018 and 2024. Patients receiving left ventricular pacing or no postprocedural temporary pacing were excluded. Outcomes included procedural complications, ICU utilization, hospital stay, permanent pacemaker (PPM) implantation and 30-day clinical events. Results Among 544 patients, 124 received a screw-in lead and 420 a conventional lead. Screw-in leads were associated with significantly lower ICU utilization after TAVR (23% vs 99%, p < 0.001). Despite longer temporary pacing duration (2.4 ± 1.4 days vs 1.2 ± 1.6 days, p < 0.001), lead dislodgement was infrequent and numerically lower with screw-in leads (0.8% vs 3.3%). Rates of major vascular complications, bleeding, tamponade, and atrial fibrillation were similar between groups. At 30 days, PPM implantation rates were comparable (50.0% vs 46.2%; p=0.455), whereas hospital stay was shorter with screw-in leads (5.2 ± 4.2 vs 7.4 ± 7.8 days, p < 0.001). Mortality and major adverse events were similar. Conclusions In TAVR recipients with pre-existing RBBB, screw-in temporary pacing leads were associated with reduced ICU utilization and shorter hospitalization without increasing complications or adverse outcomes, supporting their use for prolonged postprocedural monitoring.
No takes yet. Share an insight, caveat, or question.
Fischer et al. (2026) conducted an observational in Pre-existing right bundle branch block (RBBB) undergoing transcatheter aortic valve replacement (TAVR) (n=544). Screw-in temporary pacing leads vs. Conventional temporary pacing leads was evaluated on ICU utilization (p=<0.001). Screw-in temporary pacing leads were associated with significantly lower ICU utilization (23% vs 99%, p<0.001) and shorter hospital stay compared to conventional leads after TAVR.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: