OBJECTIVE: To identify common photon-counting CT (PCCT) virtual monochromatic imaging (VMI) settings for accurate hepatic fat fraction (FF) quantification across body sizes, including large body habitus. METHODS: ) abdomens. Phantoms were repeatedly scanned on a PCCT system (NAEOTOM Alpha) at 120 and 140 kV. CT numbers were measured on VMIs from 40-190 keV in 1-keV increments. Linear regression between ground-truth FF and Hounsfield units (HU) was used to estimate FF, and the settings minimizing relative root-mean-square error (rRMSE) in both phantoms were identified. Prediction accuracy was further assessed using leave-one-out cross-validation on the four lesions with FF of 10%-30%, with each prediction based on a regression model fitted from the remaining five lesions. Robust performance (rRMSE ≤ 5%) was maintained across 68-70 keV. Individual-lesion relative errors ranged from - 3.9% to 5.8%, while cross-validated prediction relative errors for FF 10%-30% ranged from - 5.59% to 5.98%. Performance improved over previously reported single- and dual-energy CT results, while using 40% less dose. CONCLUSION: PCCT provided consistent hepatic FF quantification across phantom sizes. A single 140-kV protocol with 68-70-keV VMIs yielded low errors, supporting potential opportunistic liver fat assessment across body sizes, including large body habitus.