Key result
QDOT MICRO catheter ablation achieves 100% acute pulmonary vein isolation in paroxysmal AF.
Why the study?
A novel QDOT MICRO catheter with optimized temperature control and microelectrodes was designed to streamline ablation workflow for paroxysmal atrial fibrillation, requiring evaluation of its workflow, performance, and safety.
Temperature-controlled QDOT MICRO ablation supports feasible PVI workflow in paroxysmal AF; leaves open long-term durability and comparative effectiveness.
BACKGROUND: A novel QDOT MICRO (Biosense Webster, Inc., Irvine, CA) catheter with optimized temperature control and microelectrodes was designed to incorporate real-time temperature sensing with contact force detection and microelectrodes to streamline ablation workflow. The QDOT-MICRO feasibility study evaluated the workflow, performance, and safety of temperature-controlled catheter ablation in patients with symptomatic paroxysmal atrial fibrillation with conventional ablation setting. METHODS: This was a non-randomized, single-arm, first-in-human study. The primary outcome was pulmonary vein isolation (PVI), confirmed by entrance block after adenosine and/or isoproterenol challenge. Safety outcomes included incidences of early-onset primary adverse events (AEs) and serious adverse device effects (SADEs). Device performance was evaluated via physician survey. RESULTS: All evaluated patients (n = 42) displayed 100% PVI. Two primary AEs (4.8%) were reported: 1 pericarditis and 1 vascular pseudoaneurysm. An additional SADE of localized infection was reported in 1 patient. No stroke, patient deaths, or other unanticipated AEs were reported. Average power delivered was 32.1±4.1 W, with a mean temperature of 40.8°C±1.6°C. Mean procedure (including 20-minute wait), fluoroscopy, and radiofrequency application times were 129.8, 6.7, and 34.0 minutes, respectively. On device performance, physicians reported overall satisfactory performance with the new catheter, with highest scores for satisfaction and usefulness of the temperature indicator. CONCLUSIONS: Initial clinical experience with the novel catheter showed 100% acute PVI success and acceptable safety and device performance in temperature-controlled ablation mode. There were no deaths, stroke, or unanticipated AEs. Fluoroscopy and procedural times were short and similar or better than reported for prior generation catheters.
No takes yet. Share an insight, caveat, or question.
A 2020 study studied symptomatic paroxysmal atrial fibrillation (n=42). QDOT MICRO temperature-controlled catheter ablation was evaluated on pulmonary vein isolation (PVI), confirmed by entrance block after adenosine and/or isoproterenol challenge. Temperature-controlled catheter ablation using the novel QDOT MICRO catheter achieved 100% acute pulmonary vein isolation in patients with paroxysmal atrial fibrillation.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: