In a multimode photoacoustic cell, the acoustic response depends on the spatial distribution of absorbed power. For heat deposition confined to a short axial region, its position is represented by the power-weighted centroid. We investigate whether cavity asymmetry improves localisation using microphone amplitude and phase measurements, which would let a sensor monitor beam displacement from its own acoustic signal. A finite-element model of a dumbbell cell predicts that unequal cavities move the microphone ratio away from unity. We printed a symmetric and an asymmetric cell, placed a microphone in each cavity, and translated a compact optically heated absorber along the axis. In the asymmetric cell the ratio at the centre rose to 2.84, close to the predicted 2.82. Within-scan validation recovered the stage-referenced axial position with a mean absolute error of 1.98 mm, compared with 8.65 mm for the symmetric cell, a fourfold difference under the same regression procedure. The results motivate further investigation of cavity geometry for axial localisation.
No takes yet. Share an insight, caveat, or question.
Sheikhveisi et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: