Key result
Transcatheter closure with the Amplatzer Membranous VSD Occluder achieves ~96% complete closure at 6 months.
Why the study?
Surgical repair of perimembranous ventricular septal defects carries a small but definite risk of morbidity and mortality, prompting evaluation of the initial safety and efficacy of transcatheter closure using the new Amplatzer Membranous VSD Occluder in the US.
Is transcatheter closure using the Amplatzer Membranous VSD Occluder safe and effective for closing perimembranous ventricular septal defects?
Is transcatheter closure using the Amplatzer Membranous VSD Occluder safe and effective for closing perimembranous ventricular septal defects?
Phase I results demonstrate that transcatheter closure of perimembranous VSDs using the Amplatzer Membranous VSD Occluder is technically feasible with high 6-month closure rates (96%) and acceptable safety in patients over 8 kg, supporting further clinical trials.
Should not yet change practice for perimembranous VSD closure; leaves open need for randomized safety and efficacy data.
Objectives This phase I study attempted to report the initial safety and efficacy results of transcatheter closure of perimembranous ventricular septal defects (PmVSDs) using the new Amplatzer Membranous VSD Occluder (AGA Medical Corp., Golden Valley, Minnesota) in the U.S. Background The most common congenital heart disease is PmVSD. Surgical repair is widely accepted, but still carries a small but definite risk of morbidity and mortality. Methods Between October 2003 and August 2004, a total of 35 patients with PmVSD underwent an attempt of transcatheter closure under transesophageal and/or intracardiac echocardiographic guidance. The median age was 7.7 years (range, 1.2 to 54.4 years) and median weight was 25 kg (range, 8.3 to 110 kg). The median Qp/Qs ratio was 1.8 (range, 1 to 4), and the median VSD size as assessed by echocardiography was 7 mm (range, 4 to 15 mm). Results The attempt to place a device was successful in 32 patients (91%). The median device size used was 10 mm (range, 6 to 16 mm). The complete closure rates by echocardiography at 10 min (transesophageal/intracardiac), 24 h, 1 month, and 6 months (transthoracic) were 47% (15/32), 63% (20/32), 78% (25/32), and 96% (27/28), respectively. The median fluoroscopy time was 36 min (range, 14 to 191 min), and the median total procedure time was 121 min (range, 67 to 276 min). Three patients (8.6%) had serious adverse events of complete heart block, peri-hepatic bleeding, and rupture of tricuspid valve chordae tendineae. No other patient encountered serious adverse events during the follow-up. Conclusions Transcatheter closure of a PmVSD is technically feasible and seems safe enough in children over 8 kg in weight to warrant continuation of clinical trials to assess the long-term safety and efficacy.
No takes yet. Share an insight, caveat, or question.
Fu et al. (2006) studied Perimembranous ventricular septal defects (PmVSDs) (n=35). Transcatheter closure using the Amplatzer Membranous VSD Occluder was evaluated on Complete VSD closure rate by echocardiography at 6 months. Transcatheter PmVSD closure using the Amplatzer Membranous VSD Occluder achieved 91% placement success, 96% complete closure at 6 months, and an 8.6% serious adverse event rate.