Key result
FARAWAVE linked to ~6-fold higher reporting odds of manufacturing issues versus VARIPULSE.
Why the study?
Post-marketing safety comparisons of device- and patient-related complications between commercially available pulsed-field ablation platforms remain limited.
Does the FARAWAVE pulsed-field ablation catheter have a different profile of device- and patient-related complications compared to the VARIPULSE catheter in the FDA MAUDE database?
Observational
Does the FARAWAVE pulsed-field ablation catheter have a different profile of device- and patient-related complications compared to the VARIPULSE catheter in the FDA MAUDE database?
Post-marketing FDA MAUDE surveillance reveals distinct safety profiles between commercial pulsed-field ablation catheters, with Farawave linked to more manufacturing and post-insertion device issues and Varipulse linked to more software-related and specific clinical complications.
MAUDE signals indicate device-specific complication patterns for Farawave versus Varipulse; hypothesis-generating and should not yet change practice.
Introduction Pulsed-field ablation (PFA) is a novel energy source that induces selective and irreversible electroporation of cardiomyocytes without collateral damage to surrounding tissues. Post-marketing safety comparisons between commercially available PFA platforms remain limited. Objective To compare device- and patient-related complications with the FARAWAVE and VARIPULSE PFA catheters using the FDA MAUDE database. Methods The FDA MAUDE database was searched for adverse events associated with commercial PFA catheters from January 1, 2024, to June 1, 2025. Reports were classified by manufacturer and FDA product codes and grouped into device malfunctions, patient injuries, or deaths. Disproportionality analysis was conducted using reporting odds ratios (ROR) with 95% confidence intervals (CI). Results A total of 1,480 events were reported for Farawave and 95 for Varipulse. Circulatory complications were more frequent with Varipulse (325 vs 17; ROR 1.41, 95% CI 0.83 – 2.41), and the neurologic complications (114 vs 50; ROR 0.09, 95% CI 0.06 – 0.14). Arrhythmias (112 vs 4; ROR 2.01, 95% CI 0.73 – 5.58) were similar. Life-threatening events were similar (25 vs 2; ROR 0.86, 95% CI 0.20 – 3.69). Varipulse reported more other organ damage (17 vs 4; ROR 0.29, 95% CI 0.09 – 0.87). Regarding device complications, 1488 in Farawave and 103 in Varipulse were reported. Farawave was associated with significantly higher reporting odds for manufacturing-related issues (292 vs 4; ROR 6.04, 95% CI 2.21–16.55) and defects identified during pre-insertion (122 vs 2; ROR 4.51, 95% CI 1.10–18.51). Post-insertion device-related events were also more frequently reported with Farawave (670 vs 22; ROR 3.02, 95% CI 1.86–4.88). Conversely, software/interface problems were disproportionately reported with Varipulse (26 vs 6; ROR 0.30, 95% CI 0.12–0.75 relative to Farawave). Other device-related complications were less frequently reported with Farawave (377 vs 70; ROR 0.16, 95% CI 0.10–0.25) Conclusion In this analysis, Varipulse was more frequently associated with circulatory and other organ complications, while neurologic events differed between systems. No significant differences were observed in arrhythmias or life-threatening events. Farawave was more often associated with manufacturing and post-insertion device issues, whereas Varipulse had more software-related and certain clinical complications. These differences suggest distinct safety profiles that should be evaluated in further studies.
No takes yet. Share an insight, caveat, or question.
Makdah et al. (2026) conducted an observational in Adverse events associated with pulsed-field ablation catheters. FARAWAVE pulsed-field ablation catheter vs. VARIPULSE pulsed-field ablation catheter was evaluated on Manufacturing-related device complications (among device- and patient-related complications) (ROR 6.04, 95% CI 2.21-16.55). FARAWAVE had higher reporting odds for manufacturing (ROR 6.04; 95% CI 2.21-16.55) and post-insertion issues (ROR 3.02), while VARIPULSE had more software and neurologic issues (ROR 0.09).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: