Key result
Ventricular leadless pacemaker implantation achieves ~100% procedural success with low complication rates through mid-term follow-up.
Why the study?
After more than a decade of clinical use worldwide, real-world data on the long-term performance and management of end-of-life leadless pacemakers remain limited.
Does ventricular leadless pacemaker implantation provide safe and durable technical performance, stable pacing parameters, and long-term battery longevity in routine clinical practice?
Cohort (n=447)
No
Does ventricular leadless pacemaker implantation provide safe and durable technical performance, stable pacing parameters, and long-term battery longevity in routine clinical practice?
In a 10-year single-center prospective cohort of 447 patients, ventricular leadless pacemakers demonstrated a 99.6% implantation success rate, low acute complication rates, stable long-term electrical performance, and high battery durability with only one device reaching end-of-life.
May support leadless pacemaker selection in practice; extends single-center observational data on success and durability.
Background Leadless pacemakers (L-PM) eliminate the need for transvenous leads and pacemaker pockets, reducing lead-related complications. After more than a decade of clinical use worldwide, real-world data on long-term performance and management of end-of-life devices remain limited. We report a 10-year single-center experience focusing on procedural practice evolution, long-term pacing parameters, safety and battery longevity. Methods Single-center prospective registry including all consecutive patients undergoing ventricular L-PM implantation from May 2015 to November 2025. Clinical characteristics, procedural data, pacing parameters and estimated battery longevity were collected at implant and at latest follow-up (FUP). Results The cohort included 447 patients (mean age 80.0±9.4 years; 64% male). Successful implantation occurred in 445 patients (99.6%). Main pacing indications were high-grade AV block (64%) and atrial fibrillation with pauses (27%). Notably, 8% of implants were performed after extraction of an infected conventional system and 17% following transcatheter structural heart procedures. Procedural efficiency improved over the years, with global mean procedure and fluoroscopy times of 41.1±24.7 min and 4.4±3.8 min. Acute major complications were infrequent: pericardial effusion occurred in 0.9% (one tamponade), and 0.4% required conversion to a transvenous pacemaker (one due to acute L-PM dysfunction and the other due to a small RV cavity). Additionally, two lymphatic access-site complications were reported, one requiring surgical treatment. No device dislodgment occurred. During a mean FUP of 2.5 ± 2.3 years, pacing parameters remained stable (threshold: 0.64 to 0.60V; R-wave amplitude: 12.3 to 12.5mV). Mean ventricular pacing was 59±39% and estimated battery longevity was >8 years in 84%. Among those with >5-year of FUP (N=60; mean FUP 6.7±2.4 years), 80% maintained >5 years of predicted longevity. Three upgrades to CRT or LBBAP occurred due to pacing-induced LV dysfunction. One true end-of-life event occurred 6 years post-implant (100% VP, high thresholds), successfully managed with a second L-PM. There were 119 deaths (27%), none device-related. Conclusion Over 10 years of experience, L-PM implantation proved to be a robust and safe pacing modality in routine practice, maintaining stable electrical performance and excellent battery durability, with only one device reaching end-of-life. Ongoing FUP will clarify long-term replacement strategies and lifetime device management in this growing population.
No takes yet. Share an insight, caveat, or question.
Pereira et al. (2026) conducted a cohort in Bradyarrhythmias requiring ventricular pacing (primarily high-grade AV block and atrial fibrillation with pauses) (n=447). Ventricular leadless pacemaker implantation was evaluated on Procedural success, acute major complications, long-term pacing parameters, and estimated battery longevity. Ventricular leadless pacemaker implantation achieved 99.6% procedural success with 0.9% pericardial effusion and >8 years estimated battery longevity in 84% of patients over 2.5 years mean follow-up.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: