Key result
Physiological feature auditing reveals sampling-rate-induced data leakage in 12 of 16 tested definitions.
Why the study?
The literature treats hand-crafted features computed from sampled physiological signals as if they were free of sampling-rate influence and compares them across devices, datasets, and rates on that assumption. Furthermore, the systematic effect of sampling rate is almost never entered into the uncertainty budget.
Sampling rate acts as a predictable and quantifiable influence quantity on hand-crafted physiological features—including ECG and photoplethysmography—requiring a ratio of sampling rate to content bandwidth of at least 10 (e.g., 440–520 Hz for the 40 Hz diagnostic band) and proper accounting of structural formula vulnerability when comparing features across devices and datasets.
Answers are grounded in this paper and cite their sources.
Sampling rate may bias hand-crafted features in diagnostics; leaves open rate-standardized extraction for cross-device validity.
In the vocabulary of metrology, a feature computed from a sampled signal is a measurement result, and the sampling rate is an influence quantity — a quantity that is not the measurand but affects the measured value. The literature treats hand-crafted features as if they were free of this influence and compares them across devices, datasets and rates on that assumption. We show that the assumption fails in a way that is predictable, quantifiable, and — for one large family of features — impossible to repair by sampling faster. We introduce the invariance order 𝑝 of a feature, defined by 𝐹(𝑥𝑟) = 𝐹(𝑥∞)(1 + 𝑂(𝑟 −𝑝 )) with 𝑟 = 𝑓𝑠/𝐵 the ratio of sampling rate to content bandwidth, and prove four propositions that assign 𝑝 from the algebraic form of a definition alone: sample-indexed quantities have 𝑝 = 1 and a relative bias that does not depend on 𝑟; finite difference and crossing-count functionals have 𝑝 = 2; bin-normalised spectral entropy decays only as 1/log𝑟; moment-based features have 𝑝 = ∞. A fifth proposition links the unobservable bias to the observable rank statistic through AUC = Φ(Δ/√2𝑠𝑎 2 + 𝑠𝑒 2), which explains why the AUC exponent and the bias exponent diverge once the bias saturates. In the terms of the GUM the influence produces a systematic effect that belongs in the uncertainty budget but is almost never entered there. The bias law is verified on synthetic signals with known ground truth (fitted slopes −2.27, −2.17, −2.02; 𝑅 2 ≥ 0.998) and on five signal families — 500 Hz ECG, 360 Hz ECG, 256 Hz photoplethysmography, 333 Hz radar displacement and 200 kHz acoustic emission from additive manufacturing — where the measured exponents are 2.03/1.86/1.95, 2.29/2.06/1.94, 2.41/2.37/1.96, and 1.03/1.11/0.93 for the sample-indexed controls. Fitted on 𝑟 ≤ 25 and extrapolated to 𝑟 = 100, the law predicts held-out AUCs to within 0.002. The link is verified on 211 cells across three datasets with 𝑅 2 = 0.9988, 0.9988 and 0.9987 and no free parameters. Auditing sixteen feature definitions drawn from the literature, twelve leak at 𝑟 = 2.5 with |AUC − 0.5| > 0.1. The effect sizes separate by two orders of magnitude between orders: 𝑑 = 15–7035 for sample-indexed features against 𝑑 = 0.45–1.32 for difference-based ones. A paired design across seven conditions (four ECG leads, two window lengths, two modalities) reproduces the taxonomy in 164 of 168 cells at FDR < 0.05. We give a design rule (r at least 10, i.e. 440–520 Hz for the 40 Hz diagnostic band), a static procedure that assigns an order before any experiment is run (14 of 16 definitions within 0.1), and evidence that p-values alone would misclassify a zero-effect feature as leaking. The contribution is a controlled, statically executable diagnostic, not a simulation of any particular instrument. Analog effects are excluded by design and are additive, so every number reported here is a strict lower bound on the device-level discrepancy: the audit ranks definitions by their structural vulnerability to the sampling grid, a property of the formula that holds whatever hardware is attached.
No takes yet. Share an insight, caveat, or question.
Yudong Xu (2026) studied this question. Definition-level invariance order audit of sampling rate effects vs. Sample-indexed vs difference-based feature definitions across sampling rate ratios was evaluated on Sampling-rate-induced feature bias and leakage (|AUC - 0.5| > 0.1 at r = 2.5) (d = 15-7035 vs d = 0.45-1.32, p=FDR < 0.05). Auditing 16 physiological feature definitions showed 12 leaked at r=2.5 (|AUC-0.5|>0.1), with effect sizes of d=15-7035 for sample-indexed vs d=0.45-1.32 for difference-based features.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: