Answers are grounded in this paper and cite their sources.
Sentinel lymph node mapping is clinically important, but optical localization at depth is limited by tissue scattering.This preliminary phantom study evaluates whether multispectral optoacoustic imaging can detect lymph-node-mimicking inclusions embedded in a tissue-like scattering medium.Model nodes containing indocyanine green (ICG) or ICG J-aggregates were used as near-infrared absorbing targets.Compact lymph node models were placed at controlled depths in a Lipofundin-based tissue-mimicking liquid.Two contrast formulations were compared: conventional ICG and self-assembled ICG J-aggregates with a shifted near-infrared absorption band.The measurements were designed to assess detectability, depth dependence and spectral separability under simplified phantom conditions.Optoacoustic measurements were performed with a custom 3D tomographic system based on ultrasound detection of laser-induced acoustic signals [1].Excitation wavelengths covered the main spectral features of ICG and ICG J-aggregates [2].Representative spectra and reconstructed projections of embedded lymph node models are shown in Fig. 1.The same phantom platform is compatible with complementary fluorescence readout.