Rapid and continuous monitoring of blood hematocrit (HCT) levels is critical for detecting hemoconcentration and plasma leakage associated with dengue virus progression. This paper presents a label-free, metamaterial-based microwave sensor designed to detect dielectric variations corresponding to changes in HCT. The proposed structural design utilizes a keyhole-shaped resonator. The sensor’s transmission coefficients (S21) were numerically evaluated using simulated dielectric properties representing an extended modelled HCT range of 0% to 80%. The simulated analysis demonstrates a systematic response, linear in resonance frequency (R2 = 0.996) and curved in magnitude (R2 = 0.812), achieving a resonance shift sensitivity of 4.5 MHz/%HCT and a magnitude sensitivity of 0.0599 dB/%HCT. Based on these simulations, a 20% relative rise from a patient’s baseline HCT (assuming = 50%) corresponds to a 45 MHz resonance shift and a −0.599 dB magnitude deviation. As an initial proof of concept, the prototype was tested with alcohol–water mixtures. Over 0–40% alcohol the resonance moved to lower frequencies, whereas above 60%, it reversed. This compact microwave sensor is a candidate framework for point-of-care HCT monitoring, subject to validation on blood.
No takes yet. Share an insight, caveat, or question.
Qureshi et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: