Answers are grounded in this paper and cite their sources.
We propose goal-driven self-recursive optical research in which agents use physical evidence to revise the hardware, software, models, and protocols used to pursue a scientific goal. The framework can be instantiated with different stimulus systems and measurement devices. We formalize its updates and derive a relation between physical attainability, command access, and model approximation error. We demonstrate the concept in an optical see-through color-matching case study using an iPad, RayNeo X3 glasses, and a TOP 300 probe coupled to a CAS140D spectroradiometer, both from Instrument Systems. A revised glasses application adds USB-controlled RGB holds and frame acknowledgements, keeping model-generated colors displayed during spectral acquisition. The interface passes 12 live control checks. Spectral feedback reduces one target color difference from 19.756 to 1.211 ΔE_76^*; all five targets improve in a separate comparison. These measurements demonstrate the concept in one configuration. The resulting software and protocols remain available for subsequent investigations.
No takes yet. Share an insight, caveat, or question.
Zhu et al. (2026) studied this question.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: