Key result
Balloon-sizing Peanut Sign identifies spiral ASDs to guide successful closure after device embolization.
Why the study?
Spiral atrial septal defects may complicate transcatheter closure and increase the risk of device embolisation.
Does recognition of the bilobed 'Peanut Sign' during balloon sizing facilitate appropriate device selection and stable transcatheter closure in a pediatric patient with a spiral atrial septal defect?
Case Report (n=1)
No
Does recognition of the bilobed 'Peanut Sign' during balloon sizing facilitate appropriate device selection and stable transcatheter closure in a pediatric patient with a spiral atrial septal defect?
A bilobed, double-waisted balloon deformation ('Peanut Sign') during balloon sizing can identify spiral atrial septal defects and guide device upsizing to prevent occluder embolisation.
Answers are grounded in this paper and cite their sources.
Peanut sign on balloon sizing may guide device selection after embolization in spiral ASD; leaves open prospective validation before routine use.
Spiral atrial septal defects may complicate transcatheter closure and increase the risk of device embolisation. We describe a 5-year-old boy with pulmonary stenosis and a secundum atrial septal defect in whom an initially implanted 9 mm device embolised. At repeat intervention, balloon sizing revealed a bilobed, peanut-shaped configuration with a double waist, suggestive of a spiral atrial septal defect. A 15 mm device was subsequently implanted, achieving stable closure. The "Peanut Sign" may serve as a useful marker for identifying spiral atrial septal defects and guiding appropriate device selection.
No takes yet. Share an insight, caveat, or question.
Hein et al. (2026) conducted a case report in Spiral atrial septal defect and moderate valvular pulmonary stenosis (n=1). Transcatheter atrial septal defect closure guided by balloon sizing ('Peanut Sign') using a 15 mm Occlutech Figulla Flex II occluder was evaluated on Stable transcatheter closure of spiral atrial septal defect without residual shunt or embolisation. Recognition of a bilobed 'Peanut Sign' with 14 and 15 mm waists during balloon sizing identified a spiral atrial septal defect and guided stable closure with a 15 mm occluder after a 9 mm device embolised.