Key result
Coupled microcatheter-RotaWire advancement enables successful rotational atherectomy in complex calcified LAD occlusion.
Why the study?
RotaWire delivery can be challenging in severely calcified device-uncrossable coronary lesions when conventional wire exchange through a microcatheter is not feasible.
Case Report (n=1)
Answers are grounded in this paper and cite their sources.
Coupled advancement of a dual-lumen microcatheter and RotaWire as a single unit offers a potential bailout technique when a RotaWire crosses a severely calcified lesion directly but cannot be advanced distally.
RotaWire delivery can be challenging in severely calcified device-uncrossable coronary lesions when conventional wire exchange through a microcatheter is not feasible. We report a 76-year-old man with a severely calcified mid-left anterior descending coronary artery (LAD) lesion in which neither a microcatheter nor a 0.75 mm balloon could cross. A second channel was created using a penetration wire through a dual-lumen microcatheter, allowing direct RotaWire crossing. However, the RotaWire could not be advanced sufficiently into the distal LAD. After intentional proximal pullback of the dual-lumen microcatheter, the microcatheter and RotaWire were held together and advanced as a single unit, enabling adequate distal RotaWire positioning and successful rotational atherectomy. This case highlights a distinct challenge after successful direct RotaWire crossing and demonstrates coupled advancement with a dual-lumen microcatheter as a potential bailout technique when distal RotaWire advancement remains difficult.
No takes yet. Share an insight, caveat, or question.
Kurogi et al. (2026) conducted a case report in Severely calcified device-uncrossable mid-left anterior descending coronary artery (LAD) lesion (n=1). Coupled advancement of dual-lumen microcatheter and RotaWire as a single unit following penetration wire crossing was evaluated on Adequate distal RotaWire positioning and successful rotational atherectomy. Coupled advancement of a dual-lumen microcatheter and RotaWire as a single unit enabled distal wire positioning and rotational atherectomy in 1 patient (76-year-old man) with a calcified LAD lesion.