This is the authors' abstract. We don't add key points for this paper.
BACKGROUND: Liver rupture caused by blunt trauma can result in severe haemorrhage and death. Percutaneous puncture embolisation is a potential minimally invasive approach for haemostasis, but manual puncture is limited by difficult bleeding-site localization and hand-eye coordination. METHODS: We developed a semi-automatic puncture system integrating a robotic arm and a compact rack-pinion driven end-effector. Positioning and posture accuracy were evaluated, followed by ex vivo and in vivo porcine liver embolisation experiments. RESULTS: The system can achieve mean positioning and posture accuracies of 0.36 mm and 1.03° under load-free conditions, and 1.83 mm and 2.67° in the soft-tissue phantom, respectively. Embolic material was delivered to the target region, and cessation of visible bleeding was observed within approximately 20 s in the ex vivo and in vivo experiments. CONCLUSIONS: The results support the preliminary feasibility of this approach for percutaneous puncture embolisation of complex hepatic lacerations.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Wong et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: