Does a novel 12-hole gold open-irrigation catheter improve lesion size and safety at low irrigation flow rates compared to a platinum-iridium catheter in a canine thigh muscle model?
A novel 12-hole gold irrigated-tip catheter allows for safe radiofrequency ablation at lower irrigation flow rates while creating larger lesions compared to a standard platinum-iridium catheter in an animal model.
BACKGROUND: High irrigation rates during radiofrequency (RF) ablation may cause fluid overload and limit lesion size. This in vivo animal study assessed the safety and efficacy of RF ablation at low irrigation rates using a novel 12-hole gold catheter. METHODS: A total of 103 lesions, created on the thigh of five mongrel dogs, were analyzed. Lesions were created using a 12-hole irrigated gold-tip (Au) and a six-hole irrigated platinum-iridium (PtIr) catheter (both 7F/3.5-mm electrode; BIOTRONIK SE 15 mL/min: 36.0 ± 1.5°C vs 46.9 ± 5.4°C, P < 0.0001) and perpendicular position (15 mL/min: 35.5 ± 1.2°C vs 38.4 ± 2.5°C, P = 0.003). The number of steam pops between Au and PtIr was comparable for parallel (8 mL/min: 14% vs 27%, P = 0.65; 15 mL/min: 14% vs 43%, P = 0.21) and perpendicular orientation (8 mL/min: 25% vs 17%, P = 1.00; 15 mL/min: 18% vs 0%, P = 0.48). Au created larger volumes than PtIr at 8 mL/min irrigation (861 ± 251 mm(3) vs 504 ± 212 mm(3) , P = 0.004); however, for 15 mL/min, volumes were comparable (624 ± 269 mm(3) vs 768 ± 466 mm(3) , P = 0.46). No coagulum formation was observed for any of the catheters on the surface and catheter tip. CONCLUSION: RF ablation at low flow rate using a novel 12-hole irrigation Au catheter is safe and results in larger lesions than with a PtIr electrode.