Why the study?
Does a novel 12-hole gold open-irrigation catheter improve lesion size and safety at low irrigation flow rates compared to a platinum-iridium catheter in a canine thigh muscle model?
Does a novel 12-hole gold open-irrigation catheter improve lesion size and safety at low irrigation flow rates compared to a platinum-iridium catheter in a canine thigh muscle model?
A novel 12-hole gold irrigated-tip catheter allows for safe radiofrequency ablation at lower irrigation flow rates while creating larger lesions compared to a standard platinum-iridium catheter in an animal model.
Answers are grounded in this paper and cite their sources.
May support low-flow ablation with gold-tip catheters in models; leaves open human translation and safety.
BACKGROUND: High irrigation rates during radiofrequency (RF) ablation may cause fluid overload and limit lesion size. This in vivo animal study assessed the safety and efficacy of RF ablation at low irrigation rates using a novel 12-hole gold catheter. METHODS: A total of 103 lesions, created on the thigh of five mongrel dogs, were analyzed. Lesions were created using a 12-hole irrigated gold-tip (Au) and a six-hole irrigated platinum-iridium (PtIr) catheter (both 7F/3.5-mm electrode; BIOTRONIK SE & CO, KG, Berlin, Germany) in parallel and perpendicular orientation. RF current was delivered for 60 seconds at 30 W using 8 mL/min and 15 mL/min irrigation. Electrode temperature, steam pops, lesion dimensions, and coagulum formation were recorded. RESULTS: Electrode temperatures were lower for Au compared to PtIr in parallel (8 mL/min: 38.1 ± 1.7°C vs 48.0 ± 4.8°C, P < 0.0001; 15 mL/min: 36.0 ± 1.5°C vs 46.9 ± 5.4°C, P < 0.0001) and perpendicular position (15 mL/min: 35.5 ± 1.2°C vs 38.4 ± 2.5°C, P = 0.003). The number of steam pops between Au and PtIr was comparable for parallel (8 mL/min: 14% vs 27%, P = 0.65; 15 mL/min: 14% vs 43%, P = 0.21) and perpendicular orientation (8 mL/min: 25% vs 17%, P = 1.00; 15 mL/min: 18% vs 0%, P = 0.48). Au created larger volumes than PtIr at 8 mL/min irrigation (861 ± 251 mm(3) vs 504 ± 212 mm(3) , P = 0.004); however, for 15 mL/min, volumes were comparable (624 ± 269 mm(3) vs 768 ± 466 mm(3) , P = 0.46). No coagulum formation was observed for any of the catheters on the surface and catheter tip. CONCLUSION: RF ablation at low flow rate using a novel 12-hole irrigation Au catheter is safe and results in larger lesions than with a PtIr electrode.
No takes yet. Share an insight, caveat, or question.
Akca et al. (2013) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: