Does the Omron M3 Intellisense upper arm BP monitor meet ESH-IP2 validation criteria for self-measurement in patients with stage 3-5 CKD?
The Omron M3 Intellisense automated oscillometric device is validated for accurate blood pressure self-measurement in predialysis patients with stage 3-5 chronic kidney disease.
BACKGROUND: Encouragement of home blood pressure (BP) monitoring has a great potential to improve hypertension control rates. The purpose of this study was to test validation of the Omron M3 Intellisense (HEM-7051-E) upper arm BP measuring monitor for self-measurement according to the European Society of Hypertension International Protocol revision 2010 (ESH-IP2) in stage 3-5 chronic kidney disease (CKD) patients. METHODS: 66 patients having CKD stage 3-5 were included in the study. Nine consecutive measurements were made according to the ESH-IP2 protocol. RESULTS: The Omron M3 Intellisense device fulfills the validation criteria of the ESH-IP2 for stage 3-5 CKD patients. CONCLUSION: Although arterial stiffness can affect accurate BP measurement, there are limited data regarding the use of automated oscillometric devices in CKD. To our knowledge, this is the first study investigating validation of an oscillometric device in stage 3-5 predialysis CKD patients. This study validates the Omron M3 Intellisense upper arm device for stage 3-5 CKD patients. New validation studies investigating other oscillometric sphygmomanometers for CKD patients and involvement of nephrologists in these studies have great potential to increase patient care in CKD.
Akpolat et al. (2011) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: