ABSTRACT The Kaymont‐Rotronics Hygroskop DT, the Beckman‐Sina and the Protimeter were tested for their ability to measure water activity (a w ) of five foods over the 0.11 – 0.85 a w range. Each instrument was calibrated at 30 ± 0.1°C against ten saturated salt solutions using a modification of the AOAC Method 32.004‐32.009. Regression of the reading versus true a w , as determined by the vapor pressure manometer (VPM), gave high r 2 values for the saturated salt solutions between 0.32 – 0.85 a w . An error of approximately 0.01 a w units was found using the 1 hr data but no significantly better regression line was found using the 24‐hr data as compared to I hr data. Measurement of the a w of five foods, however, gave values differing by an average of 0.051 a w units as compared to the VPM readings. This study demonstrates justification of the FDA cutoff a w value of 0.85 for low‐acid foods as a margin of safety.