Finger-toe pulse transit time calculated using the maximum of the second derivative algorithm and optimized signal processing strongly correlated with carotid-femoral measurements (r=0.81, P<0.0001).
Observational (n=167)
Does finger-toe pulse wave velocity (ft-PWV) accurately correlate with carotid-femoral pulse wave velocity (cf-PWV) for measuring arterial stiffness?
Finger-toe pulse wave velocity using the maximum of the second derivative algorithm provides a promising, simple, noninvasive alternative to carotid-femoral pulse wave velocity for assessing aortic stiffness.
Effect estimate: r = 0.81
p-value: p=<0.0001
BACKGROUND: Carotid-femoral pulse wave velocity (PWV) (cf-PWV) is the gold standard for measuring aortic stiffness. Finger-toe PWV (ft-PWV) is a simpler noninvasive method for measuring arterial stiffness. Although the validity of the method has been previously assessed, its accuracy can be improved. ft-PWV is determined on the basis of a patented height chart for the distance and the pulse transit time (PTT) between the finger and the toe pulpar arteries signals (ft-PTT). METHOD: The objective of the first study, performed in 66 patients, was to compare different algorithms (intersecting tangents, maximum of the second derivative, 10% threshold and cross-correlation) for determining the foot of the arterial pulse wave, thus the ft-PTT. The objective of the second study, performed in 101 patients, was to investigate different signal processing chains to improve the concordance of ft-PWV with the gold-standard cf-PWV. Finger-toe PWV (ft-PWV) was calculated using the four algorithms. RESULTS: The best correlations relating ft-PWV and cf-PWV, and relating ft-PTT and carotid-femoral PTT were obtained with the maximum of the second derivative algorithm PWV: r = 0.56, P < 0.0001, root mean square error (RMSE) = 0.9 m/s; PTT: r = 0.61, P < 0.001, RMSE = 12 ms. The three other algorithms showed lower correlations. The correlation between ft-PTT and carotid-femoral PTT further improved (r = 0.81, P < 0.0001, RMSE = 5.4 ms) when the maximum of the second derivative algorithm was combined with an optimized signal processing chain. CONCLUSION: Selecting the maximum of the second derivative algorithm for detecting the foot of the pressure waveform, and combining it with an optimized signal processing chain, improved the accuracy of ft-PWV measurement in the current population sample. Thus, it makes ft-PWV very promising for the simple noninvasive determination of aortic stiffness in clinical practice.
Obeid et al. (Wed,) reported a observational. Finger-toe pulse wave velocity (ft-PWV) using the maximum of the second derivative algorithm vs. Carotid-femoral pulse wave velocity (cf-PWV) was evaluated on Correlation between finger-toe and carotid-femoral pulse transit time (PTT) and pulse wave velocity (PWV) (r = 0.81, p=<0.0001). Finger-toe pulse transit time calculated using the maximum of the second derivative algorithm and optimized signal processing strongly correlated with carotid-femoral measurements (r=0.81, P<0.0001).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: