Investigator-discretionary MultiPoint Pacing did not significantly increase the non-responder to responder conversion rate compared to continued biventricular pacing (31.8% vs 33.8%, P=0.72).
RCT (n=467)
1:1
Yes
Cardiac resynchronization therapy non-responders (n=467)
MultiPoint Pacing (MPP) vs Continued biventricular pacing (Programmed according to physician's discretion)
Rate of conversion to echocardiographic response, p=0.72
Absolute Event Rate: 31.8% vs 33.8%
p-value: p=0.72
AIMS: To assess the impact of MultiPoint™ Pacing (MPP)-programmed according to the physician's discretion-in non-responders to standard biventricular pacing after 6 months. METHODS AND RESULTS: The study enrolled 1921 patients receiving a quadripolar cardiac resynchronization therapy (CRT) system capable of MPP™ therapy. A core laboratory assessed echocardiography at baseline and 6 months and defined volumetric non-response to biventricular pacing as <15% reduction in left ventricular end-systolic volume (LVESV). Clinical sites randomized patients classified as non-responders in a 1:1 ratio to receive MPP (236 patients) or continued biventricular pacing (231 patients) for an additional 6 months and evaluated rate of conversion to echocardiographic response. Baseline characteristics of both groups were comparable. No difference was observed in non-responder to responder conversion rate between MPP and biventricular pacing (31.8% and 33.8%, P = 0.72). In the MPP arm, 68 (29%) patients received MPP programmed with a wide LV electrode anatomical separation (≥30 mm) and shortest LV1-LV2 and LV2-RV timing delays (MPP-AS); 168 (71%) patients received MPP programmed with other settings (MPP-Other). MPP-AS elicited a significantly higher non-responder conversion rate compared to MPP-Other (45.6% vs. 26.2%, P = 0.006) and a trend in a higher conversion rate compared to biventricular pacing (45.6% vs. 33.8%, P = 0.10). CONCLUSIONS: After 6 months, investigator-discretionary MPP programming did not significantly increase echocardiographic response compared to biventricular pacing in CRT non-responders.
Building similarity graph...
Analyzing shared references across papers
Loading...
Christophe Leclercq
Electrophysiology
Haran Burri
Electrophysiology
Antonio Curnis
Azienda Socio Sanitaria Territoriale degli Spedali Civili di Brescia
European Heart Journal
King's College London
Université de Montréal
University of Geneva
Building similarity graph...
Analyzing shared references across papers
Loading...
Leclercq et al. (Mon,) conducted a rct in Cardiac resynchronization therapy non-responders (n=467). MultiPoint Pacing (MPP) vs. Continued biventricular pacing was evaluated on Rate of conversion to echocardiographic response (p=0.72). Investigator-discretionary MultiPoint Pacing did not significantly increase the non-responder to responder conversion rate compared to continued biventricular pacing (31.8% vs 33.8%, P=0.72).
synapsesocial.com/papers/6a14195fa4182367b1628d46 — DOI: https://doi.org/10.1093/eurheartj/ehz109
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: