Motivation: Osteoarthritis affects one-fourth of the world's population. Interest in applying quantitative MRI (qMRI) for early detection and follow-up of this disease has grown but ensuring reproducibility of the measurements across vendors remains necessary. Goal(s): To design and optimize an Open-Source Multi-Spin-Echo (MSE) pulse sequence for standardizing T2 Knee Articular Cartilage (KAC) mapping. Approach: Cramer-Rao Lower Bound (CRLB) was used to maximize sensitivity to KAC T2 values. Repeatability tests were carried out in the NIST phantom and a pilot in vivo scan was performed. Results: Phantom results support improved specificity and demonstrate increased reproducibility over time. Pilot in vivo KAC data proves feasibility. Impact: Multi-Spin Echo sequences for T2 mapping knee articular cartilage were made available as open-source vendor-agnostic tools to enable future quantitative MRI standardization studies. NIST phantom evaluation tests suggest improved precision with the proposed optimization. Pilot in vivo data demonstrates feasibility.
Fernandes et al. (Tue,) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: