BACKGROUND: Colistimethate sodium (CMS) is increasingly being administered as aerosol therapy in mechanically ventilated patients. Although vibrating-mesh nebulizers (VMNs) are commonly used, the impact of repeated high-dose CMS administration on expiratory filter loading and circuit pressure drop has not been well defined. This study evaluated circuit deposition patterns and changes in expiratory filter pressure drop according to the nebulizer position. METHODS: 5). Drug deposition was quantified by high-performance liquid chromatography. Pressure drop, defined as the difference between upstream and downstream pressure across the expiratory high-efficiency particulate air filter, was measured over repeated administrations and analyzed using a mixed repeated-measures model. RESULTS: = .01), with higher pressure drop after 14 treatments. In contrast, humidifier-inlet placement was associated with greater drug accumulation within the humidifier chamber (35.59 ± 6.21%) and a smaller increase in pressure drop. CONCLUSIONS: During repeated high-dose CMS administration, the nebulizer position influenced circuit deposition patterns and expiratory filter pressure drop. Placement at the Y-adapter was associated with greater expiratory limb deposition and increased filter pressure drop over time, whereas humidifier-inlet placement shifted the drug accumulation proximally within the circuit.