Influence of the Addition of Height-Adjustable Worktables on Airborne Particle Concentration in a Cleanroom According to ISO 14644-1

View Full Paper
PJPouya JaberiSDSimon DietzTWTorsten Wagner

Key Points

  • This research aims to evaluate how varying worktable heights influence airborne particle concentration in cleanrooms.
  • Measurements taken in ISO Class 6 and Class 8 cleanrooms
  • Particle concentrations recorded at heights of 70 to 120 cm
  • Data evaluated using standards ISO 14644-1 and ISO 21501-4
  • Particle concentrations were below permissible thresholds across all heights
  • Statistical significance observed in ISO Class 6 for particles >5 µm
  • No significant height effect was found in ISO Class 8 cleanroom

Abstract

Cleanrooms are essential environments for the production of sterile pharmaceuticals and medical devices, where airflow stability and contamination control are critical. In this study, the influence of adjustable worktable height on air cleanliness was examined to determine whether height variation affects unidirectional airflow and particle concentration. Measurements were conducted under laboratory conditions in an ISO Class 6 cleanroom and in an industrial ISO Class 8 production line. Particle concentrations were recorded at table heights between 70 and 120 cm using a calibrated particle counter in accordance with ISO 14644-1 and ISO 21501-4. The measured data were evaluated against the classification limits defined in ISO 14644-1. Across all height levels, particle concentrations remained well below the permissible thresholds. In the ISO Class 6 cleanroom, measurement height showed a statistically significant effect on particles >5 μm, though this represented only a small contribution to the overall variance. In contrast, no significant height effect was observed in the ISO Class 8 cleanroom. Observed local differences were attributed to airflow distribution and the placement of air supply in-/outlets rather than table height. These results confirm that height-adjustable worktables can be implemented without affecting cleanroom classification, provided that uniform FFU placement and furniture positioning are considered during design and qualification.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Jaberi et al. (2026) studied this question.

synapsesocial.com/papers/69926503eb1f82dc367a0d22https://doi.org/10.3390/app16041911
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Evaluating Clean Room Performance in the Pharmaceutical Industry: A Numerical and Experimental Investigation Based on ISO 14644–3 Standards2026 · 1 citations
  2. 2Computational Analysis of Particulate Dispersion and Environmental Purity in an ISO Class 5 Cleanroom2026
  3. 3A PAT-aligned framework for installing and operating particle counting systems to detect pre-limit particle-size distribution shifts in ISO-8 (non-sterile) controlled areas2026
  4. 4The research on the particle concentration distribution of directed airflow in cleanrooms for operators2024 · 5 citations
  5. 5Effects of contaminant diffusion on the fresh air inlets of a clean room workshop2024