PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
March 15, 2026The Journal of Sexual Medicine0 citations

(V-03) Surgical management of pump migration in inflatable penile prosthesis caused by coiled tubing & capsule formation

View Full Paper
ASAhmed Al Saeedi

Key Points

  • This case aims to highlight the surgical management of pump migration following inflatable penile prosthesis implantation due to coiling and capsular formation.
  • Reported a case of a patient with IPP implantation.
  • Performed surgical scrotal exploration for pump migration.
  • Excised fibrous capsule encasing the pump and freed coiled tubing.
  • Repositioned and secured the pump in a new location.
  • Revision surgery was completed successfully without intraoperative complications.
  • Post-surgery, the pump was functional and stable in position.
  • Patient reported ease of inflation and deflation with no recurrence during follow-up.

Abstract

Abstract Background Inflatable penile prosthesis (IPP) implantation is the definitive treatment for refractory erectile dysfunction, with high functional success rates. However, device-related complications such as pump malposition may impair usability and reduce patient satisfaction. Pump migration can result from factors including tubing redundancy, coiling, and capsular contracture, requiring surgical revision. Methods We report a case of a patient who underwent IPP implantation with a total corporal length of 26 cm (12 cm proximal, 14 cm distal), managed with a 22 (+ 4 extender) cm IPP. Postoperative recovery was uneventful, and initial pump positioning was appropriate. At follow-up, the pump was observed to have migrated proximally toward the penile shaft, resulting in difficulty with deflation and limited accessibility. Surgical scrotal exploration was performed. Intraoperatively, the pump was found displaced by coiled tubing and encased within a fibrous capsule. The capsule was excised, tubing was freed, and the pump was repositioned and placed through an opened dartos pouch. It was then secured in its new position with sutures, and the tubing was fixed with 3-0 PDS to prevent recurrence. Results The revision was completed successfully without intraoperative complications. Postoperatively, the pump was easily palpable, functional, and stable in position. The patient reported restored ease of inflation and deflation, with no recurrence during follow-up. Conclusion Pump migration caused by coiled tubing and capsular formation represents a correctable complication after IPP implantation. Timely revision surgery can restore function, prevent further morbidity, and maintain patient satisfaction. Relocating the pump to the dependent scrotum ensures better accessibility and long-term stability. This case highlights key intraoperative considerations in managing complex pump malposition.