February 1, 1988Journal of Electron Microscopy Technique909 citations

EELS log‐ratio technique for specimen‐thickness measurement in the TEM

View Full Paper
TMT. MalisSCShangcong ChengRER.F. Egerton

Key Points

Key points are not available for this paper at this time.

Abstract

We discuss measurement of the local thickness t of a transmission microscope specimen from the log-ratio formula t = lambda ln (It/I0) where It and I0 are the total and zero-loss areas under the electron-energy loss spectrum. We have measured the total inelastic mean free path lambda in 11 materials of varying atomic number Z and have parameterized the results in the form lambda = 106F (E0/Em)/ln (2 beta E0/Em) where F = (1 + E0/1,022)/(1 + E0/511)2, the incident energy E0 is in keV, the spectrum collection semiangle beta is in mrad, and Em = 7.6Z0.36. This formulation should allow absolute thickness to be determined to an accuracy of +/- 20% in most inorganic specimens.