A Bremsstrahlung spectrometer using k-edge and differential filters with image plate dosimeters

View Full Paper
CCC. D. ChenJKJ. A. KingMKM. H. Key

Key Points

Key points are not available for this paper at this time.

Abstract

A Bremsstrahlung spectrometer using k-edge and differential filtering has been used with image plate dosimeters to measure the x-ray fluence from short-pulse laser/target interactions. An electron spectrometer in front of the Bremsstrahlung spectrometer deflects electrons from the x-ray line of sight and simultaneously measures the electron spectrum. The response functions were modeled with the Monte Carlo code INTEGRATED TIGER SERIES 3.0 and the dosimeters calibrated with radioactive sources. An electron distribution with a slope temperature of 1.3 MeV is inferred from the Bremsstrahlung spectra.