March 26, 2010Journal of Cardiovascular Electrophysiology264 citations

Area Under the Real‐Time Contact Force Curve (Force–Time Integral) Predicts Radiofrequency Lesion Size in an In Vitro Contractile Model

View Full Paper
DSDipen ShahHLHendrik LambertHNHiroshi Nakagawa

Key Result

In an in vitro contractile model, the contact force-time integral correlated linearly with radiofrequency lesion volume (P<0.0001 for 20 and 40 W).

Structured PICO

Does contact force-time integral (FTI) predict radiofrequency lesion size in an in vitro contractile model?

P
Population
In vitro contractile bench model simulating the beating heart using bovine skeletal muscle
I
Intervention
Radiofrequency (RF) energy delivery (20 and 40 W for 60 seconds, 17 cc/min irrigation) with variable contact (20 g peak, 10 g nadir) and intermittent contact (20 g peak, 0 g nadir with loss of contact) at 50 and 100 movements/min
C
Comparator
Constant contact at 20 g
O
Outcome
Lesion volume and its correlation with contact force-time integral (FTI)surrogate

In an in vitro contractile model, radiofrequency lesion size correlates linearly with the contact force-time integral, demonstrating that constant contact produces larger lesions than intermittent contact despite identical peak forces.

Main Result

p-value: p=<0.0001

Abstract

FTI Predicts RF Lesion Size in Contractile Model. Introduction: Electrode tissue contact, radiofrequency (RF) power and duration are major determinants of RF lesion size. Since contact forces (CF) vary in the beating heart, we evaluated contact force–time integral (FTI) as a predictor of lesion size at constant RF power in a contractile bench model simulating the beating heart. Methods and Results: An open‐tip irrigated catheter was attached to a movable mount incorporating a dynamic force sensor allowing closed loop control to achieve desired force variations between the catheter tip and bovine skeletal muscle placed on a ground plate. RF energy (20 and 40 W for 60 seconds, 17 cc/min irrigation) was delivered during (1) constant contact (C) at 20 g, (2) variable contact (V) with a 20 g peak and 10 g nadir, and (3) intermittent contact (I) with a 20 g peak and 0 g nadir with loss of contact. V and I protocols were performed at 50 and 100 catheter movements/min and 2 systole:diastole time ratios (50:50 and 30:70). The area under the CF curve was calculated as the FTI. Measured FTI was highest in C, intermediate during V and lowest during I and correlated linearly with lesion volume (P < 0.0001 for 20 and 40 W). Lesion volume was highest in group C, intermediate in V and lowest in group I (P < 0.05 for C vs V, V vs I, and C vs I). Conclusions: Lesion size correlates linearly with measured contact FTI. Constant contact produces the largest and intermittent contact the smallest lesions despite constant RF power and identical peak contact forces. (J Cardiovasc Electrophysiol, Vol. 21, pp. 1038‐1043, September 2010)

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Shah et al. (2010) studied Radiofrequency ablation (in vitro model). Radiofrequency energy delivery with variable contact force vs. Constant contact vs variable vs intermittent contact was evaluated on Lesion volume and its correlation with contact force-time integral (FTI) (p=<0.0001). In an in vitro contractile model, the contact force-time integral correlated linearly with radiofrequency lesion volume (P<0.0001 for 20 and 40 W).