Heart rate measures from the Apple Watch, Fitbit Charge HR 2, and electrocardiogram across different exercise intensities

View Full Paper
ETElizabeth ThomsonKNKayla NussACAshley Comstock

Key Result

Heart rate monitoring accuracy by the Apple Watch (RER 2.4-5.1%) and Fitbit Charge HR 2 (RER 3.9-13.5%) compared to ECG declined as exercise intensity increased.

Study Design

Type

Observational (n=30)

Structured PICO

Do heart rate measurements by the Apple Watch and Fitbit Charge HR 2 accurately reflect ECG measurements across different exercise intensities in young adults?

P
Population
30 young adults (15 females, 15 males, mean age 23.5 ± 3.0 years) completing the Bruce Protocol.
I
Intervention
Heart rate measurement using Apple Watch and Fitbit Charge HR 2 during exercise.
C
Comparator
Heart rate measurement using electrocardiogram (ECG).
O
Outcome
Accuracy of heart rate measurements across different exercise intensities, assessed by concordance correlation coefficient (CCC), relative error rates (RER), and equivalence tests.surrogate

The accuracy of real-time heart rate monitoring by the Apple Watch and Fitbit Charge HR 2 decreases as exercise intensity increases, though the Apple Watch demonstrates lower relative error rates overall.

Abstract

This study compared heart rate (HR) measurements for the Fitbit Charge HR 2 (Fitbit) and the Apple Watch devices with HR measurements for electrocardiogram (ECG). Thirty young adults (15/15 females/males, age 23.5 ± 3.0 years) completed the Bruce Protocol. HR measurements were recorded from the ECG and both devices every minute. Average HR for each participant was calculated for very light, light, moderate, vigorous and very vigorous intensities based on ECG-measured HR. A concordance correlation coefficient (CCC) was calculated to examine the strength of the relationship between ECG measured HR and HR measured by each device. Relative error rates (RER) were also calculated to indicate the difference between each device and ECG. An equivalence test was conducted to examine the equivalence of HRs measured by devices and ECG. The Apple Watch showed lower RER (2.4–5.1%) compared with the Fitbit (3.9–13.5%) for all exercise intensities. For both devices, the strongest relationship with ECG-measured HR was found for very light PA with very high CCC (>.90) and equivalence. The strength of the relationship declined as exercise intensity increased for both devices. These findings indicate that the accuracy of real-time HR monitoring by the Apple Watch and Fitbit Charge HR2 is reduced as exercise intensity increases.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Thomson et al. (2019) conducted an observational in Young adults (n=30). Apple Watch and Fitbit Charge HR 2 vs. Electrocardiogram (ECG) was evaluated on Concordance correlation coefficient (CCC) and relative error rates (RER) of heart rate measurements. Heart rate monitoring accuracy by the Apple Watch (RER 2.4-5.1%) and Fitbit Charge HR 2 (RER 3.9-13.5%) compared to ECG declined as exercise intensity increased.

synapsesocial.com/papers/6a1000f32badbc352aff2120https://doi.org/10.1080/02640414.2018.1560644