Key result
Apple Watch and Fitbit heart rate accuracy declines vs ECG as exercise intensity increases.
Why the study?
Heart rate measurements from the Apple Watch and Fitbit Charge HR 2 devices had not been directly compared with electrocardiogram across different exercise intensities.
Do heart rate measurements by the Apple Watch and Fitbit Charge HR 2 accurately reflect ECG measurements across different exercise intensities in young adults?
Observational (n=30)
Do heart rate measurements by the Apple Watch and Fitbit Charge HR 2 accurately reflect ECG measurements across different exercise intensities in young adults?
The accuracy of real-time heart rate monitoring by the Apple Watch and Fitbit Charge HR 2 decreases as exercise intensity increases, though the Apple Watch demonstrates lower relative error rates overall.
Answers are grounded in this paper and cite their sources.
Accuracy limitations at high intensities warrant caution in athletic monitoring; hypothesis-generating and leaves open prospective validation needs.
This study compared heart rate (HR) measurements for the Fitbit Charge HR 2 (Fitbit) and the Apple Watch devices with HR measurements for electrocardiogram (ECG). Thirty young adults (15/15 females/males, age 23.5 ± 3.0 years) completed the Bruce Protocol. HR measurements were recorded from the ECG and both devices every minute. Average HR for each participant was calculated for very light, light, moderate, vigorous and very vigorous intensities based on ECG-measured HR. A concordance correlation coefficient (CCC) was calculated to examine the strength of the relationship between ECG measured HR and HR measured by each device. Relative error rates (RER) were also calculated to indicate the difference between each device and ECG. An equivalence test was conducted to examine the equivalence of HRs measured by devices and ECG. The Apple Watch showed lower RER (2.4–5.1%) compared with the Fitbit (3.9–13.5%) for all exercise intensities. For both devices, the strongest relationship with ECG-measured HR was found for very light PA with very high CCC (>.90) and equivalence. The strength of the relationship declined as exercise intensity increased for both devices. These findings indicate that the accuracy of real-time HR monitoring by the Apple Watch and Fitbit Charge HR2 is reduced as exercise intensity increases.
No takes yet. Share an insight, caveat, or question.
Thomson et al. (2019) conducted an observational in Young adults (n=30). Apple Watch and Fitbit Charge HR 2 vs. Electrocardiogram (ECG) was evaluated on Concordance correlation coefficient (CCC) and relative error rates (RER) of heart rate measurements. Heart rate monitoring accuracy by the Apple Watch and Fitbit Charge compared to ECG declined as exercise intensity increased.