Key result
Apple Watch SE and Polar Verity Sense show lower HR error than other tested wearables.
Why the study?
Heart rate measurement accuracy needed to be compared among commercially available PPG-based wearable devices against an ECG-based reference across different exercise intensities.
Do commercially available PPG-based wearable devices accurately measure heart rate compared to an ECG-based reference across different exercise intensities in healthy young adults?
Cross-Sectional (n=40)
No
Do commercially available PPG-based wearable devices accurately measure heart rate compared to an ECG-based reference across different exercise intensities in healthy young adults?
The Apple Watch SE and Polar Verity Sense provide more accurate heart rate measurements during exercise compared to the Polar Pacer Pro and Galaxy Watch 6, highlighting the need for device-specific validation of wearables.
Answers are grounded in this paper and cite their sources.
PPG wearables show variable HR accuracy by intensity; leaves open reliability for clinical exercise monitoring.
[Purpose] We compared the heart rate (HR) measurement accuracy of commercially available PPG-based wearable devices against an ECG-based reference device across different exercise intensities.[Methods] Forty healthy young adults were recruited, and data from 37 participants were analyzed. HR was simultaneously measured using Polar H10 as the ECG-based reference and four PPG-based devices: Polar Verity Sense, Polar Pacer Pro, Apple Watch SE, and Galaxy Watch 6. Participants completed low-, moderate-, and high-intensity treadmill exercise. Accuracy was assessed using mean absolute error (MAE), mean absolute percentage error (MAPE), Lin’s concordance correlation coefficient (CCC), and Bland-Altman analysis.[Results] MAE and MAPE generally decreased as intensity increased. Significant differences in MAE were observed among the devices across intensities (p = 0.002). The Polar Verity Sense and Apple Watch SE showed significantly lower MAE than the Polar Pacer Pro and Galaxy Watch 6. During low-intensity exercise, the Apple Watch SE showed lower MAE than the Polar Pacer Pro and Galaxy Watch 6, whereas during high-intensity exercise, the Polar Verity Sense showed lower MAE than the Galaxy Watch 6. CCC values increased with exercise intensity and were consistently higher for the Polar Verity Sense and Apple Watch SE. Bland-Altman analysis showed narrower limits of agreement for the Apple Watch SE and Polar Verity Sense.[Conclusion] HR measurement accuracy differed substantially among the PPG-based wearable devices. The Apple Watch SE and Polar Verity Sense showed the highest agreement with the ECG-based reference, suggesting that wearable HR data should be interpreted based on device-specific validation.
No takes yet. Share an insight, caveat, or question.
Jeong et al. (2026) conducted a cross-sectional in Healthy young adults (n=40). PPG-based wearable devices (Polar Verity Sense, Polar Pacer Pro, Apple Watch SE, Galaxy Watch 6) vs. ECG-based reference device (Polar H10) was evaluated on Mean absolute error (MAE) of heart rate measurement across devices (p=0.002). The Apple Watch SE and Polar Verity Sense demonstrated significantly lower mean absolute error in heart rate measurement compared to the Polar Pacer Pro and Galaxy Watch 6 across varying exercise intensities.
Synapse has enriched 10 related papers. Consider them for comparative context: