July 7, 2003European Journal of Cardio-Thoracic Surgery55 citationsOpen Access

Self-expandable valved stent of large size: off-bypass implantation in pulmonary position

View Full Paper
JZJ ZHOU

Key Result

Off-bypass implantation of a self-expandable valved stent in the pulmonary position was feasible in an adult pig model, demonstrating adequate hemodynamic functioning with a mean peak pressure gradient of 6.33 mmHg.

Key Points

  • The aim is to assess the feasibility of off-bypass implantation of a large self-expandable valved stent in the pulmonary valve position.
  • Implemented in 6 adult pigs with a mean body weight of 55.6 kg.
  • Used a 24 French delivery system for implantation via trans-ventricular approach through median sternotomy.
  • Evaluated with intra-vascular ultrasound and Doppler echocardiography.
  • Mean peak pressure gradient across the valve was 6.33 +/- 2.8 mmHg (Doppler) and 4.5 +/- 3.1 mmHg (invasive).
  • Mean pulmonary blood flow recorded at 3.03 +/- 0.05 l/min.
  • Intra-vascular ultrasound demonstrated complete valve opening and closure.

Structured PICO

Does off-bypass implantation of a self-expandable valved stent in the pulmonary position provide adequate hemodynamic functioning in an adult pig model?

P
Population
6 adult pigs (mean body weight 55.6 kg) underwent acute evaluation of off-bypass implantation of a self-expandable valved stent in the pulmonary position.
I
Intervention
Off-bypass implantation of a self-expandable valved stent (glutaraldehyde preserved valved bovine jugular xenograft mounted in nitinol 'Z' stent) in pulmonary valve position by trans-ventricular approach through median sternotomy
O
Outcome
Feasibility of off-bypass implantation and hemodynamic functioning (peak pressure gradient, complete opening and closure)surrogate

Off-bypass implantation of a large self-expandable valved stent in the pulmonary position is feasible and provides adequate acute hemodynamic function in a porcine model.

Abstract

OBJECTIVE: To evaluate the feasibility of the off-bypass implantation of a self-expandable valved stent of large size in pulmonary position. MATERIALS AND METHODS: A glutaraldehyde preserved valved bovine jugular xenograft with internal diameter = 22 mm, mounted in two rings of nitinol 'Z' stent, expandable from 7 to 24 mm of internal diameter, was acutely evaluated in 6 adult pigs, mean body weight 55.6 kg (range 47-67 kg). Through a stent-graft delivery system (24 French) the self expandable valved stent was implanted off-bypass in pulmonary valve position by trans-ventricular approach through median sternotomy. RESULTS: The mean diameter of the main pulmonary artery measured was 21.7 +/- 1.6 mm. The mean length of the self expandable valved stent was 23.1 +/- 0.7 mm, the mean internal diameter 21.6 +/- 0.7 mm and the mean external diameter 26.3 +/- 0.7 mm. The mean peak pressure gradient recorded across the valve was 6.33 +/- 2.8 mmHg (range 4.5-9.6 mmHg) at Doppler echocardiography, and 4.5 +/- 3.1 mmHg (range 0-7 mmHg) at invasive measurement, with a pulmonary blood flow of 3.03 +/- 0.05 l/min. Intra-vascular ultrasound showed complete opening and closure of the valve (mean area reduction from 315.08 +/- 54.13 to 0 mm2). CONCLUSIONS: (a) Off-bypass implantation of self-expandable valved stent is feasible in pulmonary position; (b) off-bypass surgical approach allows for valved stent implantation of adult size with adequate hemodynamic functioning; and (c) intra-vascular ultrasound makes implantation and evaluation easy and reproducible.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

J ZHOU (2003) studied Pulmonary valve replacement (animal model) (n=6). Off-bypass implantation of a self-expandable valved stent was evaluated on Mean peak pressure gradient across the valve. Off-bypass implantation of a self-expandable valved stent in the pulmonary position was feasible in an adult pig model, demonstrating adequate hemodynamic functioning with a mean peak pressure gradient of 6.33 mmHg.

synapsesocial.com/papers/6a20a595e033bce76a911f29https://doi.org/10.1016/s1010-7940(03)00178-7