April 1, 1986Archives of Surgery30 citations

Fate of Left-Sided Cardiac Bioprosthesis Valves in Children

View Full Paper
GKGary S. Kopf

Key Result

Porcine bioprosthetic valves in children had a 94% failure rate within six years, with catastrophic hemodynamic deterioration often requiring emergency replacement.

Study Design

Type

Cohort (n=17)

Structured PICO

What is the failure rate and clinical outcome of left-sided porcine bioprosthetic valves in children?

P
Population
17 children (mean age 8.2 years) who received 18 porcine bioprosthetic valves for aortic or mitral valve replacement, followed for up to 6 years.
E
Exposure
Porcine bioprosthetic valves (11 in aortic position, 7 in mitral position)
O
Outcome
Valve failure requiring replacementhard clinical

Porcine bioprosthetic valves in left-sided positions in children have an unacceptably high failure rate (94% within 6 years) due to rapid calcification, leading to a strong recommendation against their use in this population.

Abstract

To avoid anticoagulation and minimize thromboembolic phenomena, between 1975 and 1980 we used 18 porcine bioprosthetic valves (BPVs) to replace 11 aortic and seven mitral valves in 17 children ranging from 7 to 18 years of age (mean, 8.2 years). Ten BPVs (91%) in the aortic position had to be replaced one to six years (mean, 4.2 years) after insertion. Nine of these valves developed severe calcification with leaflet immobility and severe stenosis. The tenth valve became insufficient with a disrupted cusp. Six (86%) of seven BPVs inserted in the mitral position required replacement two to four years (mean, 3.1 years) after insertion. Massive mitral regurgitation developed in three, while in the other three mitral stenosis was prominent. All explanted BPVs exhibited calcification with disruption and loss of mobility of the leaflets. Hemodynamic deterioration often occurred catastrophically, with nine patients requiring emergency valve replacement. Elective valve replacement carried no hospital mortality, whereas emergency valve replacement carried a 33% mortality. The BPV failure rate of 94% within six years leads us to recommend against the use of biologic valves in the pediatric age group in the aortic or mitral position. Bioprosthetic valve failure may occur catastrophically and replacement should be carried out early to avoid the higher operative mortality associated with emergency surgery.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Gary S. Kopf (1986) conducted a cohort in Left-sided cardiac valve disease (n=17). Porcine bioprosthetic valves was evaluated on Bioprosthetic valve failure requiring replacement. Porcine bioprosthetic valves in children had a 94% failure rate within six years, with catastrophic hemodynamic deterioration often requiring emergency replacement.

synapsesocial.com/papers/6a34b6d586769fa257ff96e1https://doi.org/10.1001/archsurg.1986.01400040126021
Ask AI
Helpful
Bookmark
Share
View Full Paper