The Microlife WatchBP Office monitor fulfilled ANSI/AAMI/ISO validation criteria in adults and children, with a mean systolic/diastolic difference of -1.0±7.0/-4.7±5.4 mmHg compared to reference.
Cross-Sectional (n=88)
Does the Microlife WatchBP Office blood pressure monitor accurately measure blood pressure in adults and children compared to reference measurements?
The Microlife WatchBP Office blood pressure monitor meets ANSI/AAMI/ISO validation standards for accuracy in both adults and children.
Mean Difference: -1
OBJECTIVE: To assess the accuracy of the professional oscillometric blood pressure (BP) monitor Microlife WatchBP Office in adults and children according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060-2:2013 standard. METHODS: Adults and children (aged 3-12 years) were recruited to fulfil the age, sex, BP and cuff distribution criteria of the ANSI/AAMI/ISO standard using the same-arm sequential BP measurement method. Three cuffs of the test device were used for arm circumferences of 14-22, 22-32 and 32-42 cm. RESULTS: A total of 115 participants were recruited and 88 were included in the analysis (51 adults and 37 children). For criterion 1, the mean±SD of the differences between the test device and reference BP was -1.0±7.0/-4.7±5.4 mmHg (systolic/diastolic). For criterion 2, the SD of the averaged BP differences between the test device and reference per participant was 5.15/4.77 mmHg (systolic/diastolic). CONCLUSION: The professional Microlife WatchBP Office BP monitor fulfilled the requirements of the ANSI/AAMI/ISO validation standard for adults and children and can be recommended for clinical use.
Κόλλιας et al. (Tue,) reported a cross-sectional. Microlife WatchBP Office vs. Reference BP measurement was evaluated on Mean difference between test device and reference BP (Criterion 1) (MD -1.0±7.0/-4.7±5.4 mmHg (systolic/diastolic)). The Microlife WatchBP Office monitor fulfilled ANSI/AAMI/ISO validation criteria in adults and children, with a mean systolic/diastolic difference of -1.0±7.0/-4.7±5.4 mmHg compared to reference.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: