Key result
The Microlife WatchBP Office monitor fulfilled ANSI/AAMI/ISO validation criteria in adults and children, with a mean systolic/diastolic difference of -1.0±7.0/-4.7±5.4 mmHg compared to reference.
Why the study?
Does the Microlife WatchBP Office blood pressure monitor accurately measure blood pressure in adults and children compared to reference measurements?
Population
88 adults and children fulfilling age, sex, BP and cuff distribution criteria of the ANSI/AAMI/ISO standard
Comparison
Blood pressure measurement using the… vs Reference blood pressure measurement using the…
Design
Other
Authors
Loading...
Supports accuracy of Microlife WatchBP Office monitor in adults and children; extends validation data to pediatric populations.
Cross-Sectional (n=88)
Does the Microlife WatchBP Office blood pressure monitor accurately measure blood pressure in adults and children compared to reference measurements?
Mean Difference: -1
The Microlife WatchBP Office blood pressure monitor meets ANSI/AAMI/ISO validation standards for accuracy in both adults and children.
Κόλλιας et al. (2017) reported a cross-sectional. Microlife WatchBP Office vs. Reference BP measurement was evaluated on Mean difference between test device and reference BP (Criterion 1) (MD -1.0±7.0/-4.7±5.4 mmHg (systolic/diastolic)). The Microlife WatchBP Office monitor fulfilled ANSI/AAMI/ISO validation criteria in adults and children, with a mean systolic/diastolic difference of -1.0±7.0/-4.7±5.4 mmHg compared to reference.