Key result
WatchBP Office ABI accurately measures systolic BP within ~1 mmHg of observer readings across arm sizes.
Why the study?
The accuracy of the WatchBP Office ABI monitor for office blood pressure measurement over a wide range of arm circumferences was evaluated according to the ANSI/AAMI/ISO 81060-2:2013 protocol.
Cross-Sectional (n=88)
Answers are grounded in this paper and cite their sources.
Validates WatchBP Office ABI accuracy across wide arm sizes; leaves open outcome data before routine clinical adoption.
The aim of this study was to determine the accuracy of the WatchBP Office ABI monitor for office blood pressure measurement over a wide range of arm circumferences using the ANSI/AAMI/ISO 81060-2:2013 protocol. The device accuracy was tested in 88 participants whose mean±SD age was 54.5±17.6 years, whose arm circumference was 30.6±8.3 cm (range: 15-46 cm), and whose entry blood pressure (BP) was 138.3±23.4 mmHg for systolic and 83.7±14.6 mmHg for diastolic BP. Four cuffs (small, standard, large, and extra-large) suitable for arm circumferences ranging from 14.0 to 52.0 cm were used. The mean device-observer difference in the 264 separate BP data pairs was 0.7±3.8 mmHg for systolic BP and was 0.0±3.7 mmHg for diastolic BP. These data were in agreement with criterion 1 of the ANSI/AAMI/ISO 81060-2:2013 standard requirements (≤5±8 mmHg). Moreover, criterion 2 was satisfied, the mean±SD device-observer difference of the 88 participants being 0.7±3.1 and 0.0±3.2 mmHg, respectively, for systolic and diastolic BP. Good agreement between observer and device was present across the whole range of arm circumferences. These data show that the Microlife WatchBP Office ABI monitor satisfied the ANSI/AAMI/ISO 81060-2:2013 standard requirements across a wide range of arm sizes.
No takes yet. Share an insight, caveat, or question.
Palatini et al. (2018) conducted a cross-sectional in Blood pressure measurement (n=88). WatchBP Office ABI monitor vs. Observer was evaluated on Mean device-observer difference for systolic BP (MD 0.7). The WatchBP Office ABI monitor accurately measured blood pressure across a wide range of arm sizes, showing a mean device-observer difference of 0.7±3.8 mmHg for systolic BP.
Synapse has enriched 10 related papers. Consider them for comparative context: