Key result
ARM-30F-XL blood pressure monitor meets ISO accuracy standards vs. standard mercury sphygmomanometry.
Why the study?
The accuracy and safety of the ARM-30F-XL arm blood pressure monitor across multiple cuff sizes for arm circumferences of 22–52 cm required validation according to the International Standard ISO 81060-2:2018+Amd 2:2024.
Does the ARM-30F-XL arm blood pressure monitor accurately measure blood pressure compared to a mercury sphygmomanometer in individuals with an arm circumference of 22-52 cm?
Observational (n=97)
Does the ARM-30F-XL arm blood pressure monitor accurately measure blood pressure compared to a mercury sphygmomanometer in individuals with an arm circumference of 22-52 cm?
The ARM-30F-XL arm blood pressure monitor meets international standards for accuracy and is recommended for clinical and home use in individuals with arm circumferences of 22-52 cm.
Answers are grounded in this paper and cite their sources.
Supports wide-range cuff validation in large arms; leaves open outcome trials before routine adoption.
To evaluate the overall effectiveness and safety of the ARM-30F-XL arm blood pressure (BP) monitor in individuals with an arm circumference of 22–52 cm. Subjects were recruited from patients, patient companions, and individuals attending routine health examinations at the hospital. The test device was the ARM-30F-XL, and the reference device was a mercury sphygmomanometer. Using BP readings obtained from the mercury sphygmomanometer as the reference standard, the accuracy of the noninvasive BP module of the test device was evaluated to determine whether it met the specified requirements. This study analyzed valid BP data from 97 subjects. According to criterion 1, the mean difference in SBP between the test device and the reference device was 1.69 mmHg, and the SD was 5.00 mmHg. The mean difference in DBP was −2.16 mmHg, and the SD was 4.45 mmHg. The mean difference of both SBP and DBP was less than ±5 mmHg, and the SD was less than 8 mmHg, which met the requirements. According to criterion 2, the SD of SBP was 4.00 mmHg, which was less than 6.73 mmHg and met the requirements. The SD of DBP was 3.73 mmHg, which was less than 6.58 mmHg and met the requirements. The ARM-30F-XL complies completely with International Standard ISO 81060-2:2018+Amd 2:2024 and is recommended for BP measurement in individuals with an arm circumference of 22–52 cm for home and clinical use.
No takes yet. Share an insight, caveat, or question.
Qiu et al. (2026) reported an observational. ARM-30F-XL arm blood pressure monitor vs. Mercury sphygmomanometer was evaluated on Mean difference in systolic and diastolic blood pressure according to ISO 81060-2:2018+Amd 2:2024 criteria (MD 1.69 mmHg for SBP, -2.16 mmHg for DBP). The ARM-30F-XL blood pressure monitor met ISO accuracy requirements compared to a mercury sphygmomanometer, with a mean difference in SBP of 1.69 mmHg and DBP of -2.16 mmHg.