Validation of the DBP-1333b upper-arm blood pressure monitor according to the AAMI/ESH/ISO universal standard (ISO 81060-2:2018+Amd.1:2020)

View Full Paper
Ask AI
Bookmark
Share

Key result

The DBP-1333b upper-arm blood pressure monitor met AAMI/ESH/ISO standards with a mean systolic difference of 0.19 mmHg (SD 7.45) and diastolic difference of -0.59 mmHg (SD 6.47).

Why the study?

To evaluate the accuracy of the DBP-1333b upper-arm blood pressure measuring device in the adult population according to the AAMI/ESH/ISO universal standard (ISO 81060-2:2018+Amd.1:2020).

Does the DBP-1333b upper-arm blood pressure monitor accurately measure blood pressure in adults compared to a reference desktop sphygmomanometer?

Population

90 individuals in the adult population

Comparison

DBP-1333b upper-arm BP device vs desktop sphygmomanometer (XJ11D)

Design

Validation study

Authors

GHGuocun HouSuzhou Municipal HospitalYWYulu WuSichuan UniversityJWJiaying WangHong Kong Polytechnic University

Discussion

Loading...

Member takes

Implication

Offers validated upper-arm monitor option for clinicians; leaves open need for outcome data in diverse populations.

Study Design

Type

Cross-Sectional (n=90)

Structured PICO

Does the DBP-1333b upper-arm blood pressure monitor accurately measure blood pressure in adults compared to a reference desktop sphygmomanometer?

P
Population
90 adults evaluated to validate the accuracy of the DBP-1333b upper-arm blood pressure monitor against a reference desktop sphygmomanometer.
E
Exposure
DBP-1333b upper-arm blood pressure measuring device
C
Comparator
Desktop sphygmomanometer (XJ11D)
O
Outcome
Accuracy of blood pressure measurement according to AAMI/ESH/ISO universal standard (ISO 81060-2:2018+Amd.1:2020)surrogate

Main Result

Mean Difference: 0.19

The DBP-1333b upper-arm blood pressure monitor meets the AAMI/ESH/ISO universal standard for accuracy and can be recommended for use in adults.

Cite This Study

Hou et al. (2024) conducted a cross-sectional in Blood pressure measurement (n=90). DBP-1333b upper-arm blood pressure monitor vs. XJ11D desktop sphygmomanometer was evaluated on Mean difference and standard deviation of systolic and diastolic blood pressure (MD 0.19 mmHg for SBP). The DBP-1333b upper-arm blood pressure monitor met AAMI/ESH/ISO standards with a mean systolic difference of 0.19 mmHg (SD 7.45) and diastolic difference of -0.59 mmHg (SD 6.47).

synapsesocial.com/papers/6a63f5f47ff024ad283857adhttps://doi.org/10.1097/mbp.0000000000000688
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1[PP.27.03] VALIDATION OF BLOOD PRESSURE MONITORS USING THE AAMI AND ISO PROTOCOLS2016 · 8 citations
  2. 2Accuracy of ambulatory blood pressure monitors2013 · 45 citations
  3. 3A Universal Standard for the Validation of Blood Pressure Measuring Devices2018 · 468 citations
  4. 4The pursuit of accurate blood pressure measurement: A 35‐year travail2017 · 39 citations
  5. 5The German Hypertension League (Deutsche Hochdruckliga) Quality Seal Protocol for blood pressure-measuring devices2016 · 33 citations