Key result
The DBP-1333b upper-arm blood pressure monitor met AAMI/ESH/ISO standards with a mean systolic difference of 0.19 mmHg (SD 7.45) and diastolic difference of -0.59 mmHg (SD 6.47).
Why the study?
To evaluate the accuracy of the DBP-1333b upper-arm blood pressure measuring device in the adult population according to the AAMI/ESH/ISO universal standard (ISO 81060-2:2018+Amd.1:2020).
Does the DBP-1333b upper-arm blood pressure monitor accurately measure blood pressure in adults compared to a reference desktop sphygmomanometer?
Population
90 individuals in the adult population
Comparison
DBP-1333b upper-arm BP device vs desktop sphygmomanometer (XJ11D)
Design
Validation study
Authors
Loading...
Offers validated upper-arm monitor option for clinicians; leaves open need for outcome data in diverse populations.
Cross-Sectional (n=90)
Does the DBP-1333b upper-arm blood pressure monitor accurately measure blood pressure in adults compared to a reference desktop sphygmomanometer?
Mean Difference: 0.19
The DBP-1333b upper-arm blood pressure monitor meets the AAMI/ESH/ISO universal standard for accuracy and can be recommended for use in adults.
Hou et al. (2024) conducted a cross-sectional in Blood pressure measurement (n=90). DBP-1333b upper-arm blood pressure monitor vs. XJ11D desktop sphygmomanometer was evaluated on Mean difference and standard deviation of systolic and diastolic blood pressure (MD 0.19 mmHg for SBP). The DBP-1333b upper-arm blood pressure monitor met AAMI/ESH/ISO standards with a mean systolic difference of 0.19 mmHg (SD 7.45) and diastolic difference of -0.59 mmHg (SD 6.47).