Key result
DBP-1333b upper-arm blood pressure monitor passes ISO validation with ~0.2 mmHg mean systolic difference.
Why the study?
To evaluate the accuracy of the DBP-1333b upper-arm blood pressure measuring device in the adult population according to the AAMI/ESH/ISO universal standard (ISO 81060-2:2018+Amd.1:2020).
Does the DBP-1333b upper-arm blood pressure monitor provide accurate blood pressure measurements in an adult population compared to a reference mercury sphygmomanometer?
Cross-Sectional (n=90)
Observer-blinded
No
Does the DBP-1333b upper-arm blood pressure monitor provide accurate blood pressure measurements in an adult population compared to a reference mercury sphygmomanometer?
The DBP-1333b upper-arm blood pressure monitor meets the AAMI/ESH/ISO validation criteria for measurement accuracy in adults.
Answers are grounded in this paper and cite their sources.
Supports DBP-1333b accuracy per ISO in tested adults; leaves open broader validation before routine adoption.
To evaluate the accuracy of the DBP-1333b upper-arm blood pressure (BP) measuring device in the adult population according to the AAMI/ESH/ISO universal standard (ISO 81060-2:2018+Amd.1:2020). Subjects were recruited in the adult population. The test device was an arm-type electronic sphygmomanometer (DBP-1333b) and the reference device was a desktop sphygmomanometer (XJ11D). Using the BP data measured by the desktop sphygmomanometer as reference BP, the accuracy of the non-invasive BP module of the test device was evaluated to determine whether it met the requirements. Data from 90 individuals were analysed. According to Criterion 1, the mean difference of SBP between the test and reference device was 0.19 mmHg and the SD was 7.45 mmHg. The mean difference of DBP was -0.59 mmHg and the SD was 6.47 mmHg. The mean difference of both SBP and DBP was less than 5 mmHg, and the SD was less than 8 mmHg, which met the requirements. According to Criterion 2, SD of SBP was 5.79 mmHg, which was less than 6.95 mmHg and met the requirements. The SD of DBP was 5.58 mmHg, which was less than 6.93 mmHg and met the requirements. It was concluded that the DBP-1333b complies with the AAMI/ESH/ISO universal standard (ISO 81060-2:2018+Amd.1:2020) and can be recommended for use by the adults.
No takes yet. Share an insight, caveat, or question.
Hou et al. (2024) conducted a cross-sectional in Healthy and general adult population (n=90). DBP-1333b upper-arm blood pressure monitor vs. XJ11D desktop mercury sphygmomanometer was evaluated on Mean difference in systolic blood pressure (SBP) according to Criterion 1 of ISO 81060-2:2018 (MD 0.19 mmHg). The DBP-1333b upper-arm blood pressure monitor met the accuracy requirements of the ISO 81060-2:2018 standard with a mean systolic difference of 0.19 mmHg and diastolic difference of -0.59 mmHg.