Key result
DBP-6279B automated monitor passes universal validation standards, differing from mercury sphygmomanometry by ~1 mmHg.
Why the study?
The study aimed to validate the accuracy of the DBP-6279B automated inflationary oscillometric upper-arm blood pressure monitor according to the AAMI/ESH/ISO universal standard protocol.
Does the DBP-6279B automated blood pressure monitor provide accurate measurements compared to a standard mercury sphygmomanometer in adults and adolescents?
Cross-Sectional (n=88)
Single-blind
No
Does the DBP-6279B automated blood pressure monitor provide accurate measurements compared to a standard mercury sphygmomanometer in adults and adolescents?
The DBP-6279B monitor provides accurate blood pressure measurements during cuff inflation, satisfying universal validation standards for clinical and home use.
Answers are grounded in this paper and cite their sources.
Supports DBP-6279B validation per standards; leaves open generalizability across diverse populations and settings.
This study aimed to validate the accuracy of DBP-6279B, an automated inflationary oscillometric upper-arm blood pressure (BP) monitor, in the sitting position according to the AAMI/ESH/ISO (81060-2 : 2018 + Amd.1 : 2020) universal standard protocol. SBPs and DBPs were measured simultaneously on the same arm in 88 adults (female : male = 47 : 41) with a mean age of 56.85 years using a mercury sphygmomanometer (two observers) and a DBP-6279B device (one supervisor). The AAMI/ESH/ISO 81060-2 : 2018 and Amd.1 : 2020 universal standards for the validation of BP-measuring devices in adults and adolescents were followed. A total of 259 valid pairs of data were used in the analysis. According to Criterion 1, the mean difference of SBP between the test device (DBP-6279B) and the reference device (the mercury sphygmomanometer) was 0.75 mmHg, with a SD of 7.66 mmHg. The mean difference in DBP was 1.13 mmHg, with a SD of 6.14 mmHg. The mean difference of both SBP and DBP was less than 5 mmHg, and the SD was less than 8 mmHg, which met the requirements. According to Criterion 2, the mean difference of SBP between the test device and the reference device was 0.85 mmHg, and the SD was 6.56 mmHg, which was less than 6.88 mmHg and met the requirements. The mean difference in DBP was 1.27 mmHg, and the SD was 5.42 mmHg, which was less than 6.82 mmHg and met the requirements. DBP-6279B fulfilled the requirements of the AAMI/ESH/ISO universal standard (ISO 81060-2 : 2018 + Amd.1 : 2020); hence, it can be recommended for both clinical and self/home BP measurement in adults and adolescents.
No takes yet. Share an insight, caveat, or question.
Duan et al. (2023) conducted a cross-sectional in Hypertension / Outpatients (n=88). DBP-6279B automated inflationary oscillometric upper-arm blood pressure monitor vs. Mercury sphygmomanometer was evaluated on Mean difference of SBP and DBP (Criterion 1) (Mean difference SBP 0.75 mmHg). The DBP-6279B automated upper-arm blood pressure monitor met the AAMI/ESH/ISO universal standard criteria with a mean SBP difference of 0.75 mmHg and DBP difference of 1.13 mmHg compared to a mercury sphygmomanometer.