Key result
Automatic blood pressure monitor DBP-62F4B-P meets accuracy standards versus a mercury sphygmomanometer.
Why the study?
To validate the accuracy and safety of the DBP-62F4B-P automatic blood pressure monitor in adolescents and adults according to the AAMI/ESH/ISO standard.
Does the Arm-Type Fully Automatic Blood Pressure Monitor DBP-62F4B-P accurately measure blood pressure compared to a mercury sphygmomanometer in adolescents and adults?
Cross-Sectional (n=90)
Does the Arm-Type Fully Automatic Blood Pressure Monitor DBP-62F4B-P accurately measure blood pressure compared to a mercury sphygmomanometer in adolescents and adults?
The Arm-Type Fully Automatic Blood Pressure Monitor DBP-62F4B-P meets international accuracy standards for blood pressure measurement in adolescents and adults.
Answers are grounded in this paper and cite their sources.
Supports device accuracy in tested cohort; leaves open broader validation before routine adoption.
OBJECTIVE: To validate the accuracy and safety of the Arm-Type Fully Automatic Blood Pressure Monitor (model: DBP-62F4B-P) manufactured by JOYTECH Healthcare Co., Ltd. for blood pressure (BP) measurement in adolescents and adults with arm circumference 22.0-42.0 cm, in accordance with the Association for the Advancement of Medical Instrumentation (AAMI)/European Society of Hypertension (ESH)/International Organization for Standardization (ISO) (ISO 81060-2:2018+Amd 2:2024) international standard. METHODS: A total of 90 subjects were enrolled. Three valid datasets were collected from each subject, yielding 270 datasets for analysis. The mercury sphygmomanometer (reference device) data were used as the gold standard to evaluate whether the accuracy of the test device met the requirements. RESULTS: Based on 270 valid datasets, the mean difference between the test device and the reference device for SBP was -0.49 mmHg (SD = 7.29 mmHg), and for DBP was 1.45 mmHg (SD = 6.14 mmHg), meeting Criterion 1 requirements. Analysis based on individual averages of the 90 subjects showed a mean difference of -0.49 mmHg (SD = 5.57 mmHg) for SBP and 1.45 mmHg (SD = 5.05 mmHg) for DBP, both meeting the requirements of Criterion 2. The device maintained stable measurement performance across different arm circumferences, ages, and BP levels. CONCLUSION: The Arm-Type Fully Automatic Blood Pressure Monitor (model: DBP-62F4B-P) meets the accuracy and consistency requirements of the AAMI/ESH/ISO (ISO 81060-2:2018+Amd 2:2024) standard for SBP and DBP measurement in adolescents and adults with arm circumference 22.0-42.0 cm. It can be safely and effectively used for BP monitoring in clinical settings and, by extension, is suitable for home use based on its design features and validation performance.
No takes yet. Share an insight, caveat, or question.
Xiaojing Guo (2026) reported a cross-sectional. Arm-Type Fully Automatic Blood Pressure Monitor DBP-62F4B-P vs. Mercury sphygmomanometer was evaluated on Mean difference in systolic and diastolic blood pressure (MD -0.49 mmHg for SBP and 1.45 mmHg for DBP). The Arm-Type Fully Automatic Blood Pressure Monitor DBP-62F4B-P met accuracy standards with a mean difference of -0.49 mmHg for SBP and 1.45 mmHg for DBP compared to a mercury sphygmomanometer.