Key result
Hingmed DBP-01P monitor meets ISO accuracy standards with a ~1.3 mmHg mean systolic difference.
Why the study?
To determine the accuracy of the Hingmed DBP-01P monitor for clinical automatic blood pressure measurement according to the ISO 81060-2 2018 protocol.
Does the Hingmed DBP-01P clinical automatic blood pressure monitor provide accurate BP measurements compared to observer measurements in a general population?
Cross-Sectional (n=85)
Does the Hingmed DBP-01P clinical automatic blood pressure monitor provide accurate BP measurements compared to observer measurements in a general population?
The Hingmed DBP-01P blood pressure monitor with a wide-range cuff is accurate and meets ISO 81060-2:2018 standards for clinical use in the general population.
Answers are grounded in this paper and cite their sources.
Supports accuracy for clinical BP monitoring in general populations; extends ISO validation data for automated devices.
To determine the accuracy of the Hingmed DBP-01P monitor for clinical automatic blood pressure (BP) measurement according to the International Organization for Standardization (ISO) 81060-2 2018 protocol. The DBP-01P was tested in 85 subjects from the general population (mean age, 69.9 years; 43 men) using a wide-range cuff for arm circumferences from 17 to 42 cm. The mean device-observer difference was 1.32 ± 3.18 mmHg for SBP and 0.78 ± 3.15 mmHg for DBP. These data were in agreement with criterion 1 of the protocol standard requirements (≤5 ± 8 mmHg). Also, criterion 2 was satisfied with the SDs of the 85 participants being well below the maximum values required by the protocol (6.80 and 6.89 mmHg for SBP and DBP pressure, respectively). These data show that the Hingmed BP monitor DBP-01P satisfied the ISO 81060-2:2018 standard requirements for a general population across a wide range of arm sizes using a single wide-range cuff.
No takes yet. Share an insight, caveat, or question.
Fania et al. (2023) conducted a cross-sectional in General population (n=85). Hingmed DBP-01P clinical automatic blood pressure monitor vs. Observer measurement was evaluated on Accuracy according to ISO 81060-2:2018 protocol (mean device-observer difference) (Mean difference 1.32 ± 3.18 mmHg (SBP) and 0.78 ± 3.15 mmHg (DBP)). The Hingmed DBP-01P blood pressure monitor satisfied ISO 81060-2:2018 accuracy requirements in a general population, with a mean device-observer difference of 1.32 ± 3.18 mmHg for SBP.