Chromatographic method simplifies tungsten extraction in silicate rocks, suggesting improved isotope measurement accuracy.
A single‐column extraction chromatographic purification method for tungsten using the N‐benzoyl‐N‐phenylhydroxylamine (BPHA) resin was investigated in this study. This method greatly simplifies the W purification procedure, and only 10 ml of acid (5 ml 3 mol l ‐1 HCl‐0.1 mol l ‐1 HF and 5 ml 1 mol l ‐1 HF) is required to separate W from Ti, Zr, Hf, Ta and other matrix elements. The blank in the separation step was less than 0.1 ng, and the recovery of W was ~ 99%. Six independent digestion measurements of reference materials BHVO‐2, BCR‐2, RGM‐2 and JB‐2 yielded μ 182 W values (mean value with intermediate precision, 95% confidence level) of ‐10.3 ± 4.0 ( n = 6, 2 s ),‐1.2 ± 5.3 ( n = 6, 2 s ),‐1.4 ± 4.4 ( n = 6, 2 s ) and 0.6 ± 3.8 ( n = 6, 2 s ), respectively, which are in good agreement with the previously reported values.