Performance of neutron beam system for accelerator-based boron neutron capture therapy

View Full Paper
Ask AI
Bookmark
Share

Authors

VVVladislav VekslemanMHMark HarrisonKMKirill Martianov

Discussion

Loading...

Member takes

Overview

Observational analysis shows 96% uptime in neutron beam system for boron neutron capture therapy at Xiamen Humanity Hospital, suggesting efficient patient treatments.

Key Points

  • Overall uptime of the neutron beam system reached 96%, indicating reliable operation for patient treatment.
  • By late 2024, over 50 treatments were executed, showcasing the system's clinical utility and acceptance.
  • The NBS configuration allows for one treatment room, with overall usage at 27% capacity for clinical and experimental purposes.
  • The repeatability of beam parameters was within 2%, highlighting the precise control of the neutron beam system.

Cite This Study

Veksleman et al. (2025) studied this question.

synapsesocial.com/papers/68af495fad7bf08b1ead589ahttps://doi.org/10.36922/arnm025190022
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Neutronics Design of a Beam Shaping Assembly in an Accelerator-Based Neutron Source for the Boron Neutron Capture Therapy System at the University of Osaka2025 · 2 citations
  2. 2A Neutron Beam Shaping Assembly for Boron Neutron Capture Therapy of Superficial Tumors2024
  3. 3Preclinical Validation of the iBNCT001 Accelerator System for Boron Neutron Capture Therapy: In Vitro Efficacy, Beam Quality, and Radiation Safety Evaluation2026
  4. 4Neutron spectroscopy for Boron Neutron Capture Therapy beams characterization2024 · 1 citations
  5. 5VITA Accelerator Neutron Sources: Status and Research Results2026