Experiment demonstrates improved optical thickness measurement in Fabry-Perot etalon, indicating enhanced sensitivity and faster results.
We present a method for measuring the optical thickness of a high-finesse Fabry-Perot etalon using single-shot interferometry. By integrating the sample etalon within the laser cavity, a new light source is demonstrated that improves the sensitivity limitation of rear surface reflection signal. Comparative results with the conventional single-shot approach are also discussed. The light source achieved a sweep time of 49 µs and a tuning range of ~1.65 THz. The experimental results were obtained using a customized etalon with an approximate optical thickness of 730 µm, and the average optical thickness was measured as 704 ± 13 µm.
No takes yet. Share an insight, caveat, or question.
Shioda et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: