Answers are grounded in this paper and cite their sources.
Clinical experience highlights augmented reality's superior accuracy in pediatric craniosynostosis surgeries, indicating its potential in navigation.
Augmented reality (AR) has recently gained a reputation in surgical applications, providing real-time integration of virtual information into the surgeon's field of view. The aim of this paper was to describe the authors' clinical experience with AR using the Microsoft HoloLens 2 head-mounted display (HMD) in pediatric craniofacial surgery, particularly for correcting single-suture craniosynostosis. In this study, the authors compared AR-guided osteotomies with those guided by a traditional neurosurgical navigation system in a cohort of 10 consecutive pediatric patients. Osteotomy lines drawn under both AR and standard neurosurgical navigator guidance were measured using computer-aided design/computer-aided manufacturing templates. Accuracy was evaluated at the ± 1.5-mm and ± 1.0-mm thresholds. The findings demonstrated a statistically significant superior accuracy using AR guidance at the ± 1.0-mm level, achieving an average accuracy of 34% compared to 16% with standard navigation (p = 0.044). The results indicate that AR performs similarly to traditional navigation methods in terms of accuracy. These findings suggest that AR-based HMDs hold significant potential to be a reliable method of intraoperative navigation. Further studies are recommended to implement the application of this technology and assess long-term outcomes.
No takes yet. Share an insight, caveat, or question.
Ruggiero et al. (2025) studied this question.