Assessment shows point-of-care troponin I assay stability in the Emergency Department, indicating significant bias compared to laboratory assays.
Objectives To assess the imprecision and stability of the point-of-care troponin I assay in the Emergency Department and its correlation and bias to two central laboratory troponin I assays (Siemens Atellica and Abbott Alinity). Methods Imprecision and stability testing was performed on opportunistically selected samples using whole blood in the emergency department by non-laboratory trained personnel. Assay comparisons were undertaken on samples taken from participants of the Mersey Acute Coronary syndrome Rule Out Study. Results The coefficient of variation (95 % confidence interval), at the 99th percentile for the point-of-care assay, was 8.1 % (6.1–12.1 %) but with a wide confidence interval reflective of considerable scatter at values just below the 99th percentile. The 10 % limit of quantification was 7.5 ng/L (1.7–61.8 ng/L). All samples met the ≤2 ng/L stability criteria for a duration of 4 h and under. The point-of-care assay very strongly correlated and had a negative bias with the Siemens Atellica and Abbott Alinity assays, Pearson’s R=0.99 and 0.95, mean difference −29.7 ng/L and −13.3 ng/L respectively. Conclusions The Siemens VTLi point-of-care assay fulfils high-sensitivity criteria when operated by non-laboratory trained staff using whole blood in its intended environment. Lithium heparin samples are likely stable up to 4 h. Significant bias between the point-of-care and two central laboratory assays negates the use of these assays interchangeably.
No takes yet. Share an insight, caveat, or question.
Hatherley et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: