A method demonstrates accurate quantification of maduramicin ammonium in rat serum, indicating high sensitivity and reliability.
Objective: To establish a high performance liquid chromatography-tandem mass spectrometry (HPLC-MS/MS) method for the fast determination of maduramicin ammonium (MAD) in rat serum. Methods: In February 2024, rat serum samples were selected and directly injected after extraction and purification with methanol: acetonitrile (1: 1), separated on a C18 chromatographic column, and gradient-eluted using 0.1% formic acid aqueous solution -0.1% formic acid methanol solution as the mobile phase. Under optimized instrument conditions, electrospray positive ion multiple reaction monitoring (MRM) mode was employed for quantification using the external standard method, followed by methodological validation of the established approach. Results: The linearity of MAD in serum was good in the concentration range of 0.5-100 μg/L, and the correlation coefficient was 0.9997. The mean recoveries of MAD from spiked samples were 86.0%-109.6%, with the relative standard deviations were less than 10%. The limit of detection was 0.23 μg/L, The limit of detection was 0.75 μg/L. Conclusion: This method is high sensitive and reliable, which is suitable for the determination of MAD in in mouse serum.